FDA Enforcement Class II Terminated

ADVIA Centaur Systems BRAHMS PCT 100 test Sold only OUS; Catalog Number: 10378883.

Recall: Z-1659-2014 · Reported June 4, 2014

Enforcement

Recall Number
Z-1659-2014
Event ID
67953
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 4, 2014
Initiation Date
April 10, 2014
Classification Date
May 23, 2014
Termination Date
May 27, 2016
Address
333 Coney St, East Walpole, MA, 02032-1516, United States

Description

ADVIA Centaur Systems BRAHMS PCT 100 test Sold only OUS; Catalog Number: 10378883.

Reason

Samples containing fluorescein may show interference with the ADVIA Centaur¿ Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.

Code Info

Kit lots ending in/Exp Date: 020 11 July 2014 021 15 December 2014 022 24 February 2015 023 25 February 2015 024 03 July 2015

Distribution

Worldwide Distribution-USA (nationwide) and the countries of Afghanistan Algeria Armenia Austria Bahrain Belgium Bulgaria Burkina Faso Canary Islands Croatia Cyprus Czech Republic Denmark Estonia Finland France Germany Great Britain Greece Hungary Iceland Iran Iraq Ireland Israel Italy Jordan Kazakhstan Latvia Lebanon Lithuania Luxembourg Macedonia Malta Netherlands Norway Oman Pakistan Poland Portugal Qatar Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia Spain Sweden Switzerland Turkey Unit.Arab Emir. Vatican Yemen.

Quantity

16,302 kits