FDA Enforcement Class II Ongoing

Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 04U42-84 Software Version: All software versions Product Description: CELL-DYN Ruby

Recall: Z-1656-2025 · Reported April 30, 2025

Enforcement

Recall Number
Z-1656-2025
Event ID
96567
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Abbott Laboratories
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 30, 2025
Initiation Date
March 18, 2025
Classification Date
April 22, 2025
Address
100 Abbott Park Rd Bldg Ap8b, N/A, Abbott Park, IL, 60064-3502, United States

Description

Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 04U42-84 Software Version: All software versions Product Description: CELL-DYN Ruby

Reason

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.

Code Info

UDI-DI 00380740017170 SNs: 35939BG 35979BG 36019BG 36103BG 36115BG 36143BG 36171BG 36177BG 36361BG 36788BG 36866BG 54095BG 54431BG 54686BG 54787BG 54900BG 55634BG 55644BG 55656BG 55657BG 55659BG 55662BG 55665BG 55672BG 55673BG 55747BG 55859BG 55879BG 70021BG 70161BG 70204BG 70463BG 70713BG 70931BG 71185BG 71387BG 71388BG 71397BG 71398BG 71476BG 71503BG 71626BG 71637BG 71638BG 71710BG

Distribution

Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.

Quantity

45