FDA Enforcement Class II Ongoing

Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-10 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No

Recall: Z-1654-2025 · Reported April 30, 2025

Enforcement

Recall Number
Z-1654-2025
Event ID
96567
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Abbott Laboratories
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 30, 2025
Initiation Date
March 18, 2025
Classification Date
April 22, 2025
Address
100 Abbott Park Rd Bldg Ap8b, N/A, Abbott Park, IL, 60064-3502, United States

Description

Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-10 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No

Reason

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.

Code Info

UDI-DI 00380740017170 SNs: 35405BG 36034BG 36259BG 36326BG 36327BG 36328BG 36372BG 36545BG 36556BG 36583BG 36588BG 36589BG 54539BG 55361BG 55454BG 55520BG 55814BG 55828BG 72229BG

Distribution

Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.

Quantity

19