FDA Enforcement Class II Terminated

Advia Chemistry Assay: Direct HDL Cholesterol (D-HDL)

Recall: Z-1653-2016 · Reported May 18, 2016

Enforcement

Recall Number
Z-1653-2016
Event ID
73667
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Healthcare Diagnostics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
May 18, 2016
Initiation Date
March 17, 2016
Classification Date
May 11, 2016
Termination Date
September 25, 2017
Address
500 Gbc Dr Ms 514, PO BOX 6101, Newark, DE, 19702-2466, United States

Description

Advia Chemistry Assay: Direct HDL Cholesterol (D-HDL)

Reason

Siemens Healthcare Diagnostics has become aware of N-Acetylcysteine (NAC) and Metamizole (Dipyrone) interference with Trinder and Trinder-like reaction assays. Siemens has confirmed that falsely depressed results may occur on samples drawn from patients receiving N-Acetylcysteine (NAC) or Metamizole. NAC is the accepted antidote for acetaminophen toxicity and is justified in patients at signific

Code Info

Device Listing: Direct HDL Cholesterol: D126055; Catalog # 74073

Distribution

Nationwide: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico, St. Thomas, VI, Guam, Pacific Mariana Islands

Quantity

137422 units total