FDA Enforcement
Class II
Ongoing
FGFR RGQ RT-PCR Kit (RUO, not IVD) REF 8747010
Recall: Z-1652-2022
·
Reported September 7, 2022
Enforcement
- Recall Number
- Z-1652-2022
- Event ID
- 90691
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Qiagen Sciences LLC
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- September 7, 2022
- Initiation Date
- July 20, 2022
- Classification Date
- August 29, 2022
- Address
- 19300 Germantown Rd, Germantown, MD, 20874-1415, United States
Description
FGFR RGQ RT-PCR Kit (RUO, not IVD) REF 8747010
Reason
Increased occurrence rate of false positive results, lead to a false positive sample result that could subsequently be incorrectly reported from the laboratory. A false positive result could lead to the risk of the patient being exposed to an inappropriate or suboptimal anti-cancer treatment and/or unnecessarily experiencing side effects
Code Info
Lot number: 169047135 Exp Date: 12.06.2022 172017806 Exp. Date: 10.11.2022
Distribution
IN, NC, MN, TX Foreign: Belgium, China, Germany, Jordan, Italy, Netherlands, Malaysia, Spain, Thailand
Quantity
9 kits