Alaris Infusion Pump Module, Model 8100, serviced/repaired by The Biomed Guys, with the impacted bezel repair part, TIPA-8100-4410 / Alaris 8100 Bezel.

Recall: Z-1651-2021 · Reported June 9, 2021

Enforcement

Recall Number
Z-1651-2021
Event ID
87768
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
The Biomed Guys
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
June 9, 2021
Initiation Date
March 25, 2021
Classification Date
June 1, 2021
Address
1865 Ashland City Rd Ste H, Clarksville, TN, 37043-6455, United States

Description

Alaris Infusion Pump Module, Model 8100, serviced/repaired by The Biomed Guys, with the impacted bezel repair part, TIPA-8100-4410 / Alaris 8100 Bezel.

Reason

Bezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.

Code Info

Serial numbers of serviced devices, serviced 09/2020-02-2021: 12833065, 13187021, 13187811, 12838069, 12982509, 12837231, 12835186, 12837972, 12710593, 12776790, 13753893, 13605919, 12832544, 15763526, 14806203, 14806612, 14807070, 13738432, 15763547, 15865626, 14806472, 12796415,

Distribution

Us distribution to the states of IN and TX.

Quantity

22 units