FDA Enforcement
Class II
Ongoing
Puracol and Puracol Plus Collagen Wound Dressings: MSC8522 DRESSING,COLLAGEN,PURACOL,STRL,2" X 2" MSC8622EP DRESSING,COLLAGEN,PURACOL PLUS, 2X2.2"
Recall: Z-1650-2026
·
Reported April 1, 2026
Enforcement
- Recall Number
- Z-1650-2026
- Event ID
- 98484
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Medline Industries, LP
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 1, 2026
- Initiation Date
- February 10, 2026
- Classification Date
- March 26, 2026
- Address
- 3 Lakes Dr, Northfield, IL, 60093-2753, United States
Description
Puracol and Puracol Plus Collagen Wound Dressings: MSC8522 DRESSING,COLLAGEN,PURACOL,STRL,2" X 2" MSC8622EP DRESSING,COLLAGEN,PURACOL PLUS, 2X2.2"
Reason
Elevated endotoxin levels identified in affected lots. Issue may elicit strong inflammatory response if used, including local tissue irritation, delayed wound healing, or systemic effects such as fever and hypotension. In severe cases, septic shock may result.
Code Info
MSC8522 - Puracol UDI-DI 10080196285860 (each) 40080196285861 (case) Lot P2090 MSC8622EP - Puracol Plus UDI-DI 10080196296804 (each) 40080196296805 (case) Lot PL2107
Distribution
Domestic distribution nationwide US. International distribution to Mexico and Panama.
Quantity
106,621 US; 2,350 OUS