FDA Enforcement Class II Terminated

Maquet Getinge-BEQ-TOP 9004 1/4 Custom Pk w/RF32 & HMOD Material: 701054451R01

Recall: Z-1650-2019 · Reported June 5, 2019

Enforcement

Recall Number
Z-1650-2019
Event ID
82703
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Maquet Cardiovascular, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 5, 2019
Initiation Date
April 18, 2019
Classification Date
May 30, 2019
Termination Date
July 28, 2025
Address
45 Barbour Pond Dr, N/A, Wayne, NJ, 07470-2094, United States

Description

Maquet Getinge-BEQ-TOP 9004 1/4 Custom Pk w/RF32 & HMOD Material: 701054451R01

Reason

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Code Info

Batch Numbers: 3000050941 3000053625 3000069865 3000071831 3000080721 3000083489

Distribution

Nationwide

Quantity

N/A