Description
Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +3 Neck, Model Number 802802203; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented
Reason
The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
Code Info
GTIN 00887868376368, All lots distributed with IFU 87-6204-051-99 Rev D or prior
Distribution
Worldwide distribution - US Nationwide and the countries of Panama, Hong Kong, China.