FDA Enforcement
Class I
Terminated
Vitek 2 Gram Positive Susceptibility Test Cards (AST-P646 REF 420144), 20 cards/kit. Product Usage: The VITEK 2 Gram Positive Susceptibility Card is intended for use with VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.
Recall: Z-1647-2016
·
Reported May 18, 2016
Enforcement
- Recall Number
- Z-1647-2016
- Event ID
- 73413
- Classification
- Class I
- Status
- Terminated
- Product Type
- Devices
- Firm
- Biomerieux Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- May 18, 2016
- Initiation Date
- February 24, 2016
- Classification Date
- May 11, 2016
- Termination Date
- March 8, 2018
- Address
- 595 Anglum Rd, N/A, Hazelwood, MO, 63042-2320, United States
Description
Vitek 2 Gram Positive Susceptibility Test Cards (AST-P646 REF 420144), 20 cards/kit. Product Usage: The VITEK 2 Gram Positive Susceptibility Card is intended for use with VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents when used as instructed in the Product Information manual.
Reason
Product insert fails to identify performance limitation related to EUCAST breakpoints for Teicolplanin.
Code Info
Lots 746382820 exp. 6/7/17 and 746386520 exp. 7/14/17 .
Distribution
Foreign distribution in the countries of Czech Republic, Hong Kong (China), Hungary, Japan, Poland, South Africa, Switzerland, and Turkey.
Quantity
1,035 kits