FDA Enforcement Class I Ongoing

Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.

Recall: Z-1646-2026 · Reported April 8, 2026

Enforcement

Recall Number
Z-1646-2026
Event ID
98540
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
Philips Respironics, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 8, 2026
Initiation Date
March 2, 2026
Classification Date
April 2, 2026
Address
1001 Murry Ridge Ln, Murrysville, PA, 15668-8517, United States

Description

Philips Respironics Trilogy Evo, Software Version 1.05.15.00. Continuous home-use ventilator device.

Reason

In some situations, the Obstruction Alarm does not trigger within the timeframe dictated by the relevant standards and may take up to four breaths.

Code Info

All Model No.; All UDI; All Serial No.

Distribution

Worldwide Distribution. US Nationwide, Argentina, Austria, Australia, Bosnia and Herzegovina, Bangladesh, Belgium, Bulgaria, Bolivia, Brazil, Bahamas, Canada, Switzerland, Chile, China, Colombia, Czech Republic, Germany, Denmark, Algeria, Ecuador, Estonia, Egypt, Spain, Finland, France, Georgia, French Guiana, Gibraltar, Greece, Guatemala, Hong Kong, Croatia, Hungary, Indonesia, Ireland, Israel, India, Italy, Jersey, Jordan, Japan, Kenya, South Korea, Kuwait, Lebanon, Lithuania, Morocco, Montenegro, North Macedonia, Myanmar, Martinique, Malta, Mexico, Malaysia, Nigeria, Netherlands, Norway, Nepal, New Zealand, Oman, Panama, Peru, French Polynesia, Philippines, Pakistan, Poland, Portugal, Paraguay, Qatar, Reunion, Romania, Serbia, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Slovakia, Thailand, Tunisia, Turkey, Taiwan, Ukraine, United Arab Emirates, United Kingdom, Vietnam, Mayotte, South Africa, Zimbabwe.

Quantity

113,717 units