FDA Enforcement Class II Terminated

SOMATOM Emotion 16, the intended use of computed tomography is to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

Recall: Z-1645-2015 · Reported June 3, 2015

Enforcement

Recall Number
Z-1645-2015
Event ID
71101
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 3, 2015
Initiation Date
April 20, 2015
Classification Date
May 22, 2015
Termination Date
September 1, 2016
Address
51 Valley Stream Pkwy, Malvern, PA, 19355-1406, United States

Description

SOMATOM Emotion 16, the intended use of computed tomography is to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

Reason

Software bugs in VC20B SP0a or SP1 software versions may cause issues that could make it necessary to rescan patients. Syngo Main UI may crash if the Patient Browser is scrolled with arrow keys of the keyboard. Tomo images may freeze on the screen when zoomed in or out under the mode of CAREVision. Sporadic displaying error in WorkStream4D application. DB may lock under heavy and multitasks.

Code Info

SOMATOM Emotion 16 (2007) 10165977 with serial numbers: 86514 32490 80553 32552 32530 86406 86558 80542 80461 86405 32512 80441 86431 80467 86425 80442 86429 86557 80434 32581 32637 32597 80448 80431 32624 32469 80446 32578 80430

Distribution

Nationwide Distribution including Puerto Rico.

Quantity

75