FDA Enforcement
Class II
Terminated
Ascom Mobile Monitoring Gateway (versions 4.1.1 and 4.2.0) and Ascom Unite Connect for Nurse Call (version 2.2.0).
Recall: Z-1638-2016
·
Reported May 18, 2016
Enforcement
- Recall Number
- Z-1638-2016
- Event ID
- 73578
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Ascom US, INC.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 18, 2016
- Initiation Date
- December 10, 2015
- Classification Date
- May 10, 2016
- Termination Date
- October 13, 2017
- Address
- 300 Perimeter Park Dr Ste D, N/A, Morrisville, NC, 27560-9703, United States
Description
Ascom Mobile Monitoring Gateway (versions 4.1.1 and 4.2.0) and Ascom Unite Connect for Nurse Call (version 2.2.0).
Reason
Due to a malfunction of software, the secondary module will not automatically take over messaging if the primary module fails after a software upgrade to one of the other affected versions. Messages from GE Carescape/Nurse Call will not forward to handset until primary module is restored.
Code Info
Ascom Mobile Monitoring Gateway version 4.1.1 & 4.2.0 and Ascom Unite Connect for Nurse Call version 2.2.0
Distribution
CA, CT, FL, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MO, NJ, NM, OH, OK, PA, TN, TX, VA, WA, WI and Hawaii. Australia, Austria, Belgium, Canada, France, Germany, Netherlands, Poland, Qatar, Sweden, Switzerland, and United Arab Emirates.
Quantity
319