GE Healthcare CARESTATION 750c A1, Model/REF Number 1012-9755-000. Anesthesia system.
Enforcement
- Recall Number
- Z-1637-2025
- Event ID
- 96634
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- GE Medical Systems China Co., Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 7, 2025
- Initiation Date
- March 21, 2025
- Classification Date
- April 29, 2025
- Address
- Dev. Zone, National Hi-Tech; No. 19 Changjiang Road; Xin, Wuxi, N/A, China
Description
GE Healthcare CARESTATION 750c A1, Model/REF Number 1012-9755-000. Anesthesia system.
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
GTIN 00840682146425, Serial Numbers: SQT23330001WA, SQT23080006WA, SQT23080007WA, SQT23080008WA, SQT23080009WA, SQT23080010WA, SQT23080011WA, SQT24440011WA, SQT23150001WA, SQT23070001WA, SQT23070002WA, SQT24430006WA, SQT23200001WA, SQT23200002WA, SQT23200003WA, SQT23200004WA, SQT23200005WA, SQT24290003WA, SQT24510015WA, SQT24520016WA, SQT23360008WA, SQT23360002WA, SQT23360003WA, SQT23360004WA, SQT23360005WA, SQT23360006WA, SQT23360007WA, SQT23170001WA, SQT24430007WA, SQT24430008WA, SQT24430009WA, SQT24430010WA, SQT23190004WA, SQT24410005WA, SQT23050001WA, SQT23050002WA, SQT23050003WA, SQT23050004WA, SQT23060001WA, SQT23060002WA, SQT23060003WA, SQT23300002WA, SQT24230002WA, SQT24440012WA, SQT24440013WA, SQT24440014WA, SQT23400001WA, SQT23400002WA, SQT24022000WA, SQT23480001WA, SQT23370002WA, SQT23370007WA, SQT23370004WA, SQT23370006WA, SQT23370005WA, SQT23370001WA, SQT23370003WA, SQT23370008WA, SQT23370009WA, SQT23370010WA, SQT23370011WA, SQT23460001WA, SQT24160001WA, SQT23350001WA, SQT23420001WA, SQT23420002WA, SQT23190001WA, SQT23190002WA, SQT23190003WA, SQT23190005WA, SQT23190006WA, SQT23190007WA, SQT23190008WA, SQT23190009WA.
Worldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tonga, Trinidad Tobago, Tunisia, T¿rkiye, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.
74 units