FDA Enforcement Class II Terminated

TRIGEN INTERTAN 1.5 NAIL: (1) REF 71676515, 10 MM X 10 CM, 125 degree, LEFT, (2) REF 71676556, 11.5 MM X 36 CM, 125 degree, RIGHT, (3) REF 71676558, 11.5 MM X 38 CM, 125 degree, RIGHT, (4) REF 71676561, 11.5 MM X 42 CM, 125 degree, LEFT, Smith & Nephew, Inc. Orthopedic fixation device.

Recall: Z-1637-2014 · Reported May 28, 2014

Enforcement

Recall Number
Z-1637-2014
Event ID
68043
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Smith & Nephew Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
May 28, 2014
Initiation Date
April 4, 2014
Classification Date
May 20, 2014
Termination Date
July 1, 2017
Address
1450 E. Brooks Rd, N/A, Memphis, TN, 38116, United States

Description

TRIGEN INTERTAN 1.5 NAIL: (1) REF 71676515, 10 MM X 10 CM, 125 degree, LEFT, (2) REF 71676556, 11.5 MM X 36 CM, 125 degree, RIGHT, (3) REF 71676558, 11.5 MM X 38 CM, 125 degree, RIGHT, (4) REF 71676561, 11.5 MM X 42 CM, 125 degree, LEFT, Smith & Nephew, Inc. Orthopedic fixation device.

Reason

A manufacturing error resulted in the set screw being inserted too deep in the affected nails.

Code Info

(1) REF 71676515, Lot Number 13LT36602 (2) REF 71676556, Lot Numbers: 13LT36604, 13MT37354, 13MT37947, 14AT38705, 14AT38714 (3) REF 71676558, Lot Numbers: 13KT36282, 13MT37356, 13MT37664, 13MT37949, 14AT38187 (4) REF 71676561, Lot Numbers: 13LT36605, 13MT37358, 14AT38188

Distribution

Worldwide Distribution -- USA, France, Germany, Sweden, Switzerland, UK, Canada, Australia, Japan, China, India, Dubai, South Africa, South Korea, Mexico, and Venezuela.

Quantity

1,253 units total