FDA Enforcement Class II Ongoing

BEQ-HMO 50000-USA/ QUADROX-i Small, Oxygenator used during cardiopulmonary bypass Item no. 701067905

Recall: Z-1636-2023 · Reported June 7, 2023

Enforcement

Recall Number
Z-1636-2023
Event ID
92003
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Maquet Medical Systems USA
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
June 7, 2023
Initiation Date
May 18, 2023
Classification Date
May 26, 2023
Address
45 Barbour Pond Dr, N/A, Wayne, NJ, 07470-2094, United States

Description

BEQ-HMO 50000-USA/ QUADROX-i Small, Oxygenator used during cardiopulmonary bypass Item no. 701067905

Reason

1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.

Code Info

UDI-DI: 04058863019079 All lots

Distribution

Domestic distribution nationwide.

Quantity

1,784 (US)