FDA Enforcement Class II Terminated

Stryker Orthopaedics Triathlon femoral component. Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. For use in total reconstructive replacement of the knee joint

Recall: Z-1635-2015 · Reported May 27, 2015

Enforcement

Recall Number
Z-1635-2015
Event ID
70994
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Howmedica Osteonics Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
May 27, 2015
Initiation Date
August 15, 2014
Classification Date
May 19, 2015
Termination Date
January 5, 2018
Address
325 Corporate Dr, N/A, Mahwah, NJ, 07430-2006, United States

Description

Stryker Orthopaedics Triathlon femoral component. Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. For use in total reconstructive replacement of the knee joint

Reason

The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.

Code Info

Item number Lot number 5512-F-201 Lots JYET, KFVZ, KGLX ; 5512-F-202 Lots IPPOP, JYKX, KHZL, KJDN, KJHI, KJKY; 5512-F-301 Lots JTJA, KKAE, KKNB; 5512-F-302 Lots JXXF, KEYD, KFBY, KFTZ, KFXT, KHAB; 5512-F-401 Lots JWGY, KIED; 5512-F-402 Lots IUAJ, JJHT, JOMB, JWPF, JXXM, KGPE, KJMO, KLUP; 5512-F-501 Lots JXMV, KGKP; 5512-F-502 Lots KHWY, KJBK; 5512-F-601 Lots JIML, JIRX, KFAS, KIIY; 5512-F-602 Lots KEXE, KFOS, KFPI, KJAG, KKJW; 5512-F-701 Lot JKTF; 5512-F-801 Lots JWNL, JWPY; 5515-F-301 Lots IVXZD, JEPOD, JHZRA, JITNA, JITND,JJAFD, JJNYA, JJNYD, JKEND, JSNRD, JXBXA, JYFTD, JYGID, JYMTD,JYPJD, KFWHA, KFXID, KGAXD, KGJTD, KGXNA, KIBTA, KIEBA, KIGRD, KIYUD, KJABD, KJISD, KJRID; 5515-F-302 Lots IOLSD, IOLTD, IPRVD, IPTFD, IVGVA, IVLVD, JGIXD, JIGSD, JIYVD, JJSYA, JJWNA, JNVTA, JOWJD, JPFDA, JPHRA, JPIWA, JSZGD, JVYDD, JWPRA, JXBKD, JXOBA, JXXZD, JYBRA, JYBRD, JYKFD, JYKTA, JYMXA, JYNIA, JYNID, JYUED, JZDWD, JZEHD, JZFDD, KFNEA, KFOBA, KFOBD, KFOHD, KFVND, KGMFA, KGMFD, KGWGA, KGZFA, KHBGA, KIEXD, KJBFD, KJDLD, KJFXA, KJFXD, KJNPA, KKJOD; 5515-F-401 Lots JPDRA, JPIXD, JVLND, JVMTD, JXUDD, JYIDD, JYMUD, JZDSA, KABYA, KFDKD, KFMLA, KFPNA, KGAVD, KHEAA, KHKSA, KHZHA, KIATA, KIBVD, KIFMD, KIFOA, KIIFA, KJKBD, KKFYD; 5515-F-402 Lots JIFRD, JIGXD, JIMUD, JIRZD, JKESD, JLVSA, JLWTA, JOUGD, JPFEA, JPGKD, JYEIA, KAJID, KFSND, KGFLD, KGRBA, KGSHD, KGYOD, KJXDD, KKVHA, KLBHD, KLLBD; 5515-F-501 Lots HWHLA, IEXUA, IFDEA, IKEMD, JIAID, JILXA, JJBFD, JJTVD, JKGXD, JOXUA, JPMVD, JZRRA, KFSRA, KGIVD, KGKZA, KGSID, KGVHD, KGVPD, KHGPA, KHKAD, KHWRD, KHXWA, KJGBA, KKHFA, KKHFD, KLOSA; 5515-F-502 Lots IKRKA, JIDLA, JIEZD, JIVTA,JIYYD, JJWSA, JKAAD, JNVJA, JOPDD, JPFBD, JVUAD, JXBYD, JXFLA, JXTRD, JXYSA, JZHAD, JZKTD, KFNGD, KGTLA, KGTMA, KGYFA, KHGVA, KHGVA, KHGVD, KHOFD, KHPLA, KHRTA, KHWMA, KHWVA, KIDPA, KIJDA, KJGLD, KKMDD, KKOOD, KKPKA, KKWAA, KLBZA, KLMBA, KLSFA, KLSFD; 5515-F-601 Lots IPFTA, JIUMD, JJHZA, JMBTA, JWFHD, JWKTA, JXWXA, KABRD, KAGLA, KFISA, KFJXA, KGBVA, KGTDA, KHFSA, KHGNA, KHJAA, KHLIA, KHSWA, KHUAA, KHYED, KHZYA, KIGND, KJKOA; 5515-F-602 Lots JIYHD, JIZOA, JJBWD, JJWPD, JKDPD, JKERA, JPPKD, JJPSXD, JRBRA, JYTBD, KEZND, KFKDA, KFTMD, KFURA, KGIRA, KGNPD, KGRYD, KHJTD, KHPZA, KKXKA, KLKSA.

Distribution

Worldwide Distribution: US (nationwide) including PR and countries of: Australia, Canada, China, Brazil, France, Hong Kong, India, Italy, Mexico, Netherlands, Singapore, South Africa, Sweden, and Spain.

Quantity

1,147 in total