FDA Enforcement
Class II
Terminated
CDI Blood Parameter Monitoring System; Catalog Numbers or Part Numbers - 145891, 500AVHCT; Serial Numbers - BPM SN B015552 (installed into CDI 500 Monitor SN 3535)
Recall: Z-1634-2014
·
Reported May 28, 2014
Enforcement
- Recall Number
- Z-1634-2014
- Event ID
- 68240
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Terumo Cardiovascular Systems Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Telephone
- Report Date
- May 28, 2014
- Initiation Date
- May 9, 2014
- Classification Date
- May 19, 2014
- Termination Date
- July 11, 2014
- Address
- 6200 Jackson Rd, N/A, Ann Arbor, MI, 48103-9586, United States
Description
CDI Blood Parameter Monitoring System; Catalog Numbers or Part Numbers - 145891, 500AVHCT; Serial Numbers - BPM SN B015552 (installed into CDI 500 Monitor SN 3535)
Reason
There is a potential for inaccuracies in blood gas readings/measurements. There is a potential risk of respiratory alkalosis or cerebral vasoconstriction with low paCO2. If the Hemoglobin and Hematocrit measurements are not accurate, there is a potential risk of unnecesary blood transfusion with resulting risk of infection or inflammatory reaction. BPM shipped without meeting manufacturing accep
Code Info
Catalog Number or Part Number 145891 500AVHCT Serial Number: BPM SN 8015552 (installed into CDI 500 Monitor SN 3535)
Distribution
one customer in IA
Quantity
1