FDA Enforcement Class II Terminated

Stryker Howmedica Osteonics RESTORATION Acetabular Augment System; Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. Hip prosthesis component. The RESTORATION Acetabular Wedge Augments provide multiple options to address the wide range of bone deficiencies encountered in acetabular revision. The augments provide support for the shell in the acetabulum with superior and/or posterior defects.

Recall: Z-1633-2015 · Reported May 27, 2015

Enforcement

Recall Number
Z-1633-2015
Event ID
70994
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Howmedica Osteonics Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
May 27, 2015
Initiation Date
August 15, 2014
Classification Date
May 19, 2015
Termination Date
January 5, 2018
Address
325 Corporate Dr, N/A, Mahwah, NJ, 07430-2006, United States

Description

Stryker Howmedica Osteonics RESTORATION Acetabular Augment System; Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. Hip prosthesis component. The RESTORATION Acetabular Wedge Augments provide multiple options to address the wide range of bone deficiencies encountered in acetabular revision. The augments provide support for the shell in the acetabulum with superior and/or posterior defects.

Reason

The peel strength of the inner blister may have been below internal validated requirements. The strength of the seal is directly related to long term shelf life of the sterile barrier.

Code Info

Ref #:Lot Number - 5096-4615:MMR8M8; 5096-4615:MMRLYD; 5096-4615:MNAYT7; 5096-5015:MMPO6K; 5096-5015:MMRWOJ; 5096-5015:MNADVA; 5096-5815:MNA40L; 5096-5820:MMR6HX; 5096-5825:MMRLD7; 5096-6225:MMPWO5; 5096-6225:MMRM2Y; 5096-6625:MMNM07

Distribution

Worldwide Distribution: US (nationwide) including PR and countries of: Australia, Canada, China, Brazil, France, Hong Kong, India, Italy, Mexico, Netherlands, Singapore, South Africa, Sweden, and Spain.

Quantity

1,147 in total