FDA Enforcement Class I Ongoing

GE Healthcare Carestation 650 A1, Model/REF Number 1012-9650-200. Anesthesia system.

Recall: Z-1631-2025 · Reported May 7, 2025

Enforcement

Recall Number
Z-1631-2025
Event ID
96634
Classification
Class I
Status
Ongoing
Product Type
Devices
Firm
GE Medical Systems China Co., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 7, 2025
Initiation Date
March 21, 2025
Classification Date
April 29, 2025
Address
Dev. Zone, National Hi-Tech; No. 19 Changjiang Road; Xin, Wuxi, N/A, China

Description

GE Healthcare Carestation 650 A1, Model/REF Number 1012-9650-200. Anesthesia system.

Reason

Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

Code Info

GTIN 00195278439529, Serial Numbers: SM724450001MD, SM724450002MD, SM724450003MD, SM724450004MD, SM724450005MD, SM724450006MD, SM724450007MD, SM724450008MD, SM724450009MD, SM724450010MD, SM724450011MD, SM724450012MD, SM724450013MD, SM724450005WD, SM724450006WD, SM724450007WD, SM724450008WD, SM724450009WD, SM724450010WD, SM724450011WD, SM724450012WD, SM724450017WD, SM725030187WD, SM725030188WD, SM724320011WD, SM724320012WD, SM724320013WD, SM725030189WD, SM725030190WD, SM723290013WD, SM725020154WD, SM725020153WD, SM724390001WD, SM724390002WD, SM724390003WD, SM724390004WD, SM724360007WD, SM724360008WD, SM724360009WD, SM724510130WD, SM724510131WD, SM725030196WD, SM725030197WD, SM725030198WD, SM725020152WD, SM724490117WD, SM725010144WD, SM724190036WD, SM724190031WD, SM724190032WD, SM724190033WD, SM724190034WD, SM724190035WD, SM724190027WD, SM724190028WD, SM724190029WD, SM724190030WD, SM723290017WD, SM724190038WD, SM724200003WD, SM724200001WD, SM724190039WD, SM724200002WD, SM724190037WD, SM724460035WD, SM724450001WD, SM724330006WD, SM724320001WD, SM724320016WD, SM724320017WD, SM724320018WD, SM724320019WD, SM724320020WD, SM724320021WD, SM724320022WD, SM724320023WD, SM724320024WD, SM724320025WD, SM724320026WD, SM724320027WD, SM724320028WD, SM724320029WD, SM724320030WD, SM724220002WD, SM724220003WD, SM724220004WD, SM724220005WD, SM724220006WD, SM724220008WD, SM724220007WD, SM724220010WD, SM724180014WD, SM724200001MD, SM724200002MD, SM724200003MD, SM724210001MD, SM724210002MD, SM724210003MD, SM724210004MD, SM724210005MD, SM724410002MD, SM724270002WD, SM724450049WD, SM724450050WD, SM724450051WD, SM724450052WD, SM724470056WD, SM724470057WD, SM724470059WD, SM724470060WD, SM724470061WD, SM724470062WD, SM724470063WD, SM724470064WD, SM724470065WD, SM724470066WD, SM723300005MD, SM723300006MD, SM724480020MD, SM724480021MD, SM724480022MD, SM724480023MD, SM724480024MD, SM724480025MD, SM724480026MD, SM724480027MD, SM724480028MD, SM724480029MD, SM724480030MD, SM724480031MD, SM724480032MD, SM725020068MD, SM725020069MD, SM725020070MD, SM725020071MD, SM725020072MD, SM725020078MD, SM725020079MD, SM725020080MD, SM725020081MD, SM725020082MD, SM724460024WD, SM724320004WD, SM723290002WD, SM723290003WD, SM724460032WD, SM724470071WD, SM724150002WD, SM724460013WD, SM724520143WD, SM724460053WD, SM724470096WD, SM724510135WD, SM724490115WD, SM724360003WD, SM724360004WD, SM724360005WD, SM724360006WD, SM724360016WD, SM724410003WD, SM724460038WD, SM724460039WD, SM724460040WD, SM724460041WD, SM724460042WD, SM724460043WD, SM724460044WD, SM724460045WD, SM724460046WD, SM724460047WD, SM724460048WD, SM724460049WD, SM724460050WD, SM724460051WD, SM724460052WD, SM724500124WD, SM724510134WD, SM724520141WD, SM724520142WD, SM725020166WD, SM724150001WD, SM724340016WD, SM724340017WD, SM725040138MD, SM725040139MD, SM725040141MD, SM725040140MD, SM725030210WD, SM724450014MD, SM724450015MD, SM724450016MD, SM724450017MD, SM724450018MD, SM725030119MD, SM725030120MD, SM725030121MD, SM725030122MD, SM725030123MD, SM725030124MD, SM725030125MD, SM725030126MD, SM725030127MD, SM725030128MD, SM725030129MD, SM725030130MD, SM725030131MD, SM725030132MD, SM725030133MD, SM725050142MD, SM725050143MD, SM725050144MD, SM725050145MD, SM725050146MD, SM725040216WD, SM725040217WD, SM725040218WD, SM725040219WD, SM725040220WD, SM725040221WD, SM725040222WD, SM725040223WD, SM725040224WD, SM725040225WD, SM725040226WD, SM725040227WD, SM725040228WD, SM725040229WD, SM724460008WD, SM724460009WD, SM724460010WD, SM724470075WD, SM724460011WD, SM724470069WD, SM724470070WD, SM724510132WD, SM724510133WD, SM724330005WD, SM724200004WD, SM724320038WD, SM724320039WD, SM724290015WD, SM724320048WD, SM724320049WD, SM724210013WD, SM724320045WD, SM724460017WD, SM724220001WD, SM724460001WD, SM724460036WD, SM724170007WD, SM724170008WD, SM724470072WD, SM724470073WD, SM724170009WD, SM724270004WD, SM724320002WD, SM724320003WD, SM724340018WD, SM724340013WD, SM724410018WD, SM724330010WD, SM724410004WD, SM724410005WD, SM724170011WD, SM724390016WD, SM724390017WD, SM724330008WD, SM725020177WD, SM724340022WD, SM724320043WD, SM724460019WD, SM724320041WD, SM724320042WD, SM724410002WD, SM724460002WD, SM724330001WD, SM724330002WD, SM724330003WD, SM724330007WD, SM724220009WD, SM724290027WD, SM724290028WD, SM724270003WD, SM724340011WD, SM724170010WD, SM724210018WD, SM724460021WD, SM724460022WD, SM724340020WD, SM724340021WD, SM724340010WD, SM724210012WD, SM724190024WD, SM724190025WD, SM725020161WD, SM725020162WD, SM724340012WD, SM724290001WD, SM724290002WD, SM724210011WD, SM724460003WD, SM724330011WD, SM724330012WD, SM724470074WD, SM724330016WD, SM724460054WD, SM724350002WD, SM724350003WD, SM724410001WD, SM724370001WD, SM724450002WD, SM724450003WD, SM724450004WD, SM724290025WD, SM724290026WD, SM724170004WD, SM724170005WD, SM724290024WD, SM724170012WD, SM724320010WD, SM724460007WD, SM724500128WD, SM724410006WD, SM724410007WD, SM724350004WD, SM724190015WD, SM724190018WD, SM724320006WD, SM724320007WD, SM724320008WD, SM724370003WD, SM724170006WD, SM724210016WD, SM724470077WD, SM724470078WD, SM724190020WD, SM724190021WD, SM724170013WD, SM724210017WD, SM724320014WD, SM724460018WD, SM724500125WD, SM724370002WD, SM724340023WD, SM724340024WD, SM725070238WD, SM725070239WD, SM725070240WD, SM724480099WD, SM724510129WD, SM725010149WD, SM724490116WD, SM725070242WD, SM724210010WD, SM723300001MD, SM723300002MD, SM725050152MD, SM724390005WD, SM724390006WD, SM724390007WD, SM724390008WD, SM724390009WD, SM724390010WD, SM724390011WD, SM724390012WD, SM724390013WD, SM724390014WD, SM724500033MD, SM724500034MD, SM725030093MD, SM724500120WD, SM724500121WD, SM724500122WD, SM724500123WD, SM725030211WD, SM725030212WD, SM725020151WD, SM724360011WD, SM724340009WD, SM724360010WD, SM724340014WD, SM724340015WD, SM724330004WD, SM724470076WD, SM724360002WD, SM724370004WD, SM724290004WD, SM724290005WD, SM724480100WD, SM724480101WD, SM724480102WD, SM724480103WD, SM724480105WD, SM724480106WD, SM724320050WD, SM723290016WD, SM724270001WD, SM724360001WD, SM724360012WD, SM724360013WD, SM724360014WD, SM724360015WD, SM724460012WD, SM724470092WD, SM724470093WD, SM724480104WD, SM724380001WD, SM724380002WD, SM724470067WD, SM724470068WD, SM724330015WD, SM724150052MD, SM724150053MD, SM724150054MD, SM724150051MD, SM724150055MD, SM724150056MD, SM724150058MD, SM724150059MD, SM724150060MD, SM724190001MD, SM724190002MD, SM724190003MD, SM724190004MD, SM724190005MD, SM724190006MD, SM724190007MD, SM724190008MD, SM724190009MD, SM724190010MD, SM724190011MD, SM724150057MD, SM724290019WD, SM724290020WD, SM724290021WD, SM724290022WD, SM724290023WD, SM724460025WD, SM724460026WD, SM724460027WD, SM724460028WD, SM724460029WD, SM724460030WD, SM724460031WD, SM724320032WD, SM724320033WD, SM724320034WD, SM724320035WD, SM724320036WD, SM724320037WD, SM724490118WD, SM724490119WD, SM724510138WD, SM724510139WD, SM724510140WD, SM725030201WD, SM724470089WD, SM724460033WD, SM725070241WD, SM724460055WD, SM724510136WD, SM724510137WD, SM724460037WD, SM724460034WD, SM724240001WD, SM724240002WD, SM724240003WD, SM724470079WD, SM724470080WD, SM724470081WD, SM724470082WD, SM724470083WD, SM724470084WD, SM724470085WD, SM724470086WD, SM724470087WD, SM724470088WD, SM724470090WD, SM724470091WD, SM724230001WD, SM724450016WD, SM724290007WD, SM724290008WD, SM724200006WD, SM724200007WD, SM724330013WD, SM724330014WD, SM724470058WD, SM724510037MD, SM724510038MD, SM724510039MD, SM724510040MD, SM724510041MD, SM724510042MD, SM724440001MD, SM724440002MD, SM725030136MD, SM725030137MD, SM724450021WD, SM724450022WD, SM724450023WD, SM724450024WD, SM724450025WD, SM724450026WD, SM724450027WD, SM724450028WD, SM724450029WD, SM724450030WD, SM724450031WD, SM724450032WD, SM724450033WD, SM724450034WD, SM724450035WD, SM724450036WD, SM724450037WD, SM724450038WD, SM724450039WD, SM724450040WD, SM723300003MD, SM723300004MD, SM723300007MD, SM723300008MD, SM723300009MD, SM724510043MD, SM724510044MD, SM724510045MD, SM724510046MD, SM724510047MD, SM724510048MD, SM724510049MD, SM724510050MD, SM724510051MD, SM724510052MD, SM724510053MD, SM724510054MD, SM724510055MD, SM724510056MD, SM724510057MD, SM724510058MD, SM724510059MD, SM724510060MD, SM724510061MD, SM724510062MD, SM724510063MD, SM724510064MD, SM724510065MD, SM724510066MD, SM724510067MD, SM725020083MD, SM725020084MD, SM725020085MD, SM725020086MD, SM725020087MD, SM725020088MD, SM725020089MD, SM725020090MD, SM725020091MD, SM725020092MD, SM725030094MD, SM725030095MD, SM725030096MD, SM725030097MD, SM725030098MD, SM725030099MD, SM725030100MD, SM725030101MD, SM725030102MD, SM725030103MD, SM725030104MD, SM725030105MD, SM725030106MD, SM725030107MD, SM725030108MD, SM725030109MD, SM725030110MD, SM725030111MD, SM725030112MD, SM725030113MD, SM725030114MD, SM725030115MD, SM725030116MD, SM725030117MD, SM725030118MD, SM725030134MD, SM725030135MD, SM724190044WD, SM724210008WD, SM724210009WD, SM723290014WD, SM723290015WD, SM724340003WD, SM724340004WD, SM724340005WD, SM724340006WD, SM724340007WD, SM724380003WD, SM724380004WD, SM724410008WD, SM724410009WD, SM724410010WD, SM724410011WD, SM724410012WD, SM724410013WD, SM724470095WD, SM724470094WD, SM725030202WD, SM725030203WD, SM725030204WD, SM725030205WD, SM725030206WD, SM725030207WD, SM725030208WD, SM725030209WD, SM724290006WD, SM724210004WD, SM724210005WD, SM724210006WD, SM724190040WD, SM724190041WD, SM724450013WD, SM724450014WD, SM724450015WD, SM724190042WD, SM724190043WD, SM723290005WD, SM723290026WD, SM724450019WD, SM724450020WD, SM724470097WD, SM724480108WD, SM724480109WD, SM724480110WD, SM724480111WD, SM724480112WD, SM724480113WD, SM724450041WD, SM724450042WD, SM724450043WD, SM724450044WD, SM724450045WD, SM724450046WD, SM724480098WD, SM724450047WD, SM724450048WD, SM724210001WD, SM724210002WD, SM724210003WD, SM724320001MD.

Distribution

Worldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tonga, Trinidad Tobago, Tunisia, T¿rkiye, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.

Quantity

648 units