FDA Enforcement
Class II
Terminated
Carter-Thomason II Port Closure System; Model Number(s) CTI-1015P, for use in endoscopic and laparoscopic surgery.
Recall: Z-1630-2015
·
Reported May 27, 2015
Enforcement
- Recall Number
- Z-1630-2015
- Event ID
- 70828
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- CooperSurgical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 27, 2015
- Initiation Date
- March 20, 2015
- Classification Date
- May 19, 2015
- Termination Date
- July 29, 2015
- Address
- 75 Corporate Dr, N/A, Trumbull, CT, 06611-1350, United States
Description
Carter-Thomason II Port Closure System; Model Number(s) CTI-1015P, for use in endoscopic and laparoscopic surgery.
Reason
The pad printing around the suture entry holes on the 15mm Suture Guide included in the kit may not have been appropriately manufactured resulting in the pad printing flaking off around the entry holes.
Code Info
Lot 174422
Distribution
US Nationwide distribution.
Quantity
430 units