Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)
Recall: Z-1629-2020
·
Reported April 15, 2020
Enforcement
- Recall Number
- Z-1629-2020
- Event ID
- 84163
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Zeiss, Carl Inc
- Voluntary / Mandated
- FDA Mandated
- Initial Notification
- Letter
- Report Date
- April 15, 2020
- Initiation Date
- October 2, 2019
- Classification Date
- April 6, 2020
- Termination Date
- July 25, 2024
- Address
- 1 Zeiss Dr, N/A, Thornwood, NY, 10594-1939, United States
Description
Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)
Reason
When the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb laser radiation. This issue arises when a ZEISS inverted stage stand (Axio Observer or Axio Vert) is used in as a component of the LSM 700, 800 and 900 system
Code Info
(LSM 700, LSM 800 and LSM 900)
Distribution
Worldwide Distribution - US Nationwide
Quantity
US: 600; Global: 1600