Homocysteine test system - Homocysteine Reagent. 05385415190; Homocysteine 100 Tests, cobas c, COBAS INTEGRA Homocysteine Enzymatic Assay is an in vitro test for the quantitative determination of total L-homocysteine in human serum and plasma on Roche/Hitachi cobas c systems.
Enforcement
- Recall Number
- Z-1622-2015
- Event ID
- 71034
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Diagnostics Operations, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 27, 2015
- Initiation Date
- March 23, 2015
- Classification Date
- May 15, 2015
- Termination Date
- January 19, 2016
- Address
- 9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States
Description
Homocysteine test system - Homocysteine Reagent. 05385415190; Homocysteine 100 Tests, cobas c, COBAS INTEGRA Homocysteine Enzymatic Assay is an in vitro test for the quantitative determination of total L-homocysteine in human serum and plasma on Roche/Hitachi cobas c systems.
Customers complained about under-recovery of non-Roche controls and discrepant low patient results with certain Homocysteine reagent lots. This negative bias could, in the worst case, lead to inaccurately low Homocysteine results. An elevated level of Homocysteine is considered an important risk factor in the assessment of peripheral vascular disease. Falsely low values could lead to a delay of di
05385415190, Homocysteine 100Tests, cobas c, COBAS INTEGRA Lot number 69781101
Nationwide Distribution-including PR and the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, , SC, TN, TX, VA, WA, and WV.
2558 units