FDA Enforcement
Class II
Ongoing
Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.
Recall: Z-1620-2024
·
Reported May 1, 2024
Enforcement
- Recall Number
- Z-1620-2024
- Event ID
- 94306
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- SEASPINE ORTHOPEDICS CORPORATION
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- May 1, 2024
- Initiation Date
- February 28, 2024
- Classification Date
- April 23, 2024
- Address
- 5770 Armada Dr, N/A, Carlsbad, CA, 92008-4608, United States
Description
Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.
Reason
Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.
Code Info
REF/UDI-DI/Lots: MX1-001010/10889981298647/ AW161319D, AW164195D, AW166994D, AW167309D, AW167422D, AW169350F, AW169351F. AMRDXMIS/10889981303297/ AMRDXMIS-001, AMRDXMIS-002, AMRDXMIS-003, AMRDXMIS-004, AMRDXMIS-005
Distribution
US: CA, FL, CO, TX
Quantity
444 Screw Heads