FDA Enforcement Class II Terminated

Karl Storz SE & Co. KG, C-MAC Video Laryngoscope MIL # 1, for the following REF numbers: 1. 8403xx series, which includes: 8403AX, AXC, BX, BXC, DXC, EXC, GXC, HX, HXK, HXP, KXC, MXC 2. 8401xx series, which includes: 8401AX, AXC, BX, BXC, DXC, GXC, HX, HXP, KXC. Product Usage: Video Laryngoscopes are used to visualize the airways and vocal cords during endotracheal intubation and for the inspection and examination of the upper respiratory tract. Video Laryngoscopes are designed for transient use in invasive procedures through a body orifice.

Recall: Z-1618-2020 · Reported April 8, 2020

Enforcement

Recall Number
Z-1618-2020
Event ID
85089
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Karl Storz Endoscopy
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
April 8, 2020
Initiation Date
December 6, 2019
Classification Date
March 31, 2020
Termination Date
July 1, 2022
Address
2151 E Grand Ave, N/A, El Segundo, CA, 90245-5017, United States

Description

Karl Storz SE & Co. KG, C-MAC Video Laryngoscope MIL # 1, for the following REF numbers: 1. 8403xx series, which includes: 8403AX, AXC, BX, BXC, DXC, EXC, GXC, HX, HXK, HXP, KXC, MXC 2. 8401xx series, which includes: 8401AX, AXC, BX, BXC, DXC, GXC, HX, HXP, KXC. Product Usage: Video Laryngoscopes are used to visualize the airways and vocal cords during endotracheal intubation and for the inspection and examination of the upper respiratory tract. Video Laryngoscopes are designed for transient use in invasive procedures through a body orifice.

Reason

The firm has become aware that there is a potential that the glass cover at the distal end of the camera/led lighting on the video laryngoscope could become loose during reprocessing or use. If this occurs during usage, it may result in a delay in procedure and/or require removal from body.

Code Info

All Lots

Distribution

Worldwide distribution - US Nationwide and country of Canada.

Quantity

27,165 scopes