FDA Enforcement Class III Ongoing

To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree and the gastrointestinal tract Models: NA-U401SX-4022, NA-U401SX-4022-A

Recall: Z-1616-2023 · Reported May 31, 2023

Enforcement

Recall Number
Z-1616-2023
Event ID
92129
Classification
Class III
Status
Ongoing
Product Type
Devices
Firm
Olympus Corporation of the Americas
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 31, 2023
Initiation Date
March 31, 2023
Classification Date
May 19, 2023
Address
3500 Corporate Pkwy, N/A, Center Valley, PA, 18034-8229, United States

Description

To be used with ultrasound endoscopes for ultrasound guided fine needle aspiration (FNA) of submucosal and extramural lesions of the tracheobronchial tree and the gastrointestinal tract Models: NA-U401SX-4022, NA-U401SX-4022-A

Reason

Product non-conformance pertaining to the angle of the needle does not meet manufacturing specifications and remained unbent, resulting to slight resistance experienced during needle advancement and retraction.

Code Info

UDI: 04953170388286 Lot numbers: KR248740 KR248746 KR252863 KR253666 KR253681 KR260854 KR260865 KR260873 KR260875 KR260876 KR260877 KR260896 KR260917 KR260937

Distribution

Natiowide Foreign: Canada, Germany

Quantity

1297 (Boxes; 5 per Box)