FDA Enforcement Class II Terminated

COULTER DxH Diluent, Catalog No. 628017, Part No. A59956. For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells.

Recall: Z-1616-2015 · Reported May 20, 2015

Enforcement

Recall Number
Z-1616-2015
Event ID
71045
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Beckman Coulter Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 20, 2015
Initiation Date
April 13, 2015
Classification Date
May 14, 2015
Termination Date
October 1, 2015
Address
250 S Kraemer Blvd, N/A, Brea, CA, 92821-6232, United States

Description

COULTER DxH Diluent, Catalog No. 628017, Part No. A59956. For use as an isotonic buffered diluent in conjunction with a cyanide-free lytic agent for counting and sizing blood cells.

Reason

Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The use of the DxH Diluent lots listed in the recall notice may eventually result in a compromised HGB chamber.

Code Info

Lot No. 3099980 - 3512190, expiration dates: 17 April 2015 - 26 August 2016.

Distribution

Worldwide Distribution -- Argentina; Brazil; Brunei Darussalam; Canada; Chile; Colombia; Hong Kong; India; Lebanon; Macao; Mexico; Panama; Philippines; Qatar; Singapore; Taiwan; Thailand; Trinidad and Tobago; United States, including Puerto Rico; and Uruguay.

Quantity

829,888 units total (659,584 units in US)