FDA Enforcement Class II Ongoing

LightStrike Blackout Curtains, Part numbers; 610-00008, 610-00009, 610-00019, 610-00020, and 610-00024, used with the LightStrike Germ-Zapping Robots Models: PXUV4D, PXUV5D, LSMUV6-FT, or LSMUV6-SL, for whole room microbial reduction.

Recall: Z-1615-2024 · Reported May 1, 2024

Enforcement

Recall Number
Z-1615-2024
Event ID
94258
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
XENEX Disinfection Services Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 1, 2024
Initiation Date
March 18, 2024
Classification Date
April 23, 2024
Address
1074 Arion Cir Ste 116, N/A, San Antonio, TX, 78216-3085, United States

Description

LightStrike Blackout Curtains, Part numbers; 610-00008, 610-00009, 610-00019, 610-00020, and 610-00024, used with the LightStrike Germ-Zapping Robots Models: PXUV4D, PXUV5D, LSMUV6-FT, or LSMUV6-SL, for whole room microbial reduction.

Reason

Blackout curtains are no longer recommended as a protective barrier against UV/Ozone exposure to a bystander standing in the same room, but on the other side of the curtain from where a microbial reduction device UV cycle is taking place. Bystanders in the room during the cycle may experience temporary skin reddening, feeling of sand in eyes, nausea, headache, and/or mucous membrane irritation.

Code Info

Robot UDI-DIs: 00814122020033, 00814122020064, 00814122020125, 00814122020132. All serial numbers.

Distribution

US Nationwide distribution in the states of AR, AZ, CA, CO, FL, GA, HI, ID, IL, KY, LA, MI, MN, MO, NC, ND, NV, OH, OK, OR, TN, UT, WA, WV, WY.

Quantity

377