FDA Enforcement Class II Ongoing

Olympus Single Use Distal Cover Model Number MAJ-2315 used with EVIS EXERA III Duodenovideoscope TJF-Q190V; Item code: Catalogue #N5786100

Recall: Z-1613-2023 · Reported May 31, 2023

Enforcement

Recall Number
Z-1613-2023
Event ID
92115
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Olympus Corporation of the Americas
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 31, 2023
Initiation Date
April 18, 2023
Classification Date
May 19, 2023
Address
3500 Corporate Pkwy, N/A, Center Valley, PA, 18034-8229, United States

Description

Olympus Single Use Distal Cover Model Number MAJ-2315 used with EVIS EXERA III Duodenovideoscope TJF-Q190V; Item code: Catalogue #N5786100

Reason

The Single Use Distal Cover MAJ-2315 may cause mucosal injury and/or become detached from the endoscope during use. Consequences of a detached cover could include the risk of aspiration, inhalation, or obstruction that will require urgent removal, and/or burns from an uncovered distal end.

Code Info

UDI-DI: 04953170403019; Item code: Catalog N5786100 Lots Lower than H2530

Distribution

US Nationwide.

Quantity

309278 pieces