SYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, when used in conjunction with Synchron LX¿ System(s), UniCel¿ DxC 600/800 System(s) and Synchron¿ Systems Multi Calibrator, is intended for the quantitative determination of lactate concentration in human plasma and cerebrospinal fluid (CSF).
Recall: Z-1613-2013
·
Reported July 3, 2013
Enforcement
- Recall Number
- Z-1613-2013
- Event ID
- 65377
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Beckman Coulter Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- July 3, 2013
- Initiation Date
- May 28, 2013
- Classification Date
- June 27, 2013
- Termination Date
- June 3, 2014
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232, United States
Description
SYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, when used in conjunction with Synchron LX¿ System(s), UniCel¿ DxC 600/800 System(s) and Synchron¿ Systems Multi Calibrator, is intended for the quantitative determination of lactate concentration in human plasma and cerebrospinal fluid (CSF).
Reason
Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions (frozen reagent). Failed calibration of LACT may cause a delay in reporting results.
Code Info
Part Number A95550, Lot Number M210077
Distribution
Worldwide Distribution - USA Nationwide and country of Canada.
Quantity
3,353 units