FDA Enforcement Class II Terminated

Brand Name: Multiday lnfusor, Seven Day lnfusor and the Basal/Bolus lnfusor: Single day Infusor 2 ml/h, Single Day INFUSOR 2 ml/h System, Half Day INFUSOR SV 5 ml/h System, Two Day INFUSOR 2 ml/h System, Two day Infusor 2 ml/h, Multiday Infusor 0.5 ml/h, Multiday INFUSOR 0.5ml/h System, Seven Day Infusor 0.5 ml/h, Seven Day INFUSOR 0.5ml/h System, Basal-Bolus Infusor-15min., Basal/Bolus INFUSOR 0.5 x 2 ml/h System with 15 Minute Lockout, Basal-Bolus Infusor Device Listing: D018045 The Multiday lnfusor. Seven Day lnfusor and the BasaVBolus lnfusor are indicated for patients requiring slow, continuous intravenous, intra-arterial, subcutaneous or epidural administration of medications.

Recall: Z-1610-2013 · Reported July 3, 2013

Enforcement

Recall Number
Z-1610-2013
Event ID
65364
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Baxter Healthcare Corp.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
July 3, 2013
Initiation Date
June 7, 2013
Classification Date
June 27, 2013
Termination Date
June 4, 2014
Address
Route 120 And Wilson Rd, Round Lake, IL, 60073-9799, United States

Description

Brand Name: Multiday lnfusor, Seven Day lnfusor and the Basal/Bolus lnfusor: Single day Infusor 2 ml/h, Single Day INFUSOR 2 ml/h System, Half Day INFUSOR SV 5 ml/h System, Two Day INFUSOR 2 ml/h System, Two day Infusor 2 ml/h, Multiday Infusor 0.5 ml/h, Multiday INFUSOR 0.5ml/h System, Seven Day Infusor 0.5 ml/h, Seven Day INFUSOR 0.5ml/h System, Basal-Bolus Infusor-15min., Basal/Bolus INFUSOR 0.5 x 2 ml/h System with 15 Minute Lockout, Basal-Bolus Infusor Device Listing: D018045 The Multiday lnfusor. Seven Day lnfusor and the BasaVBolus lnfusor are indicated for patients requiring slow, continuous intravenous, intra-arterial, subcutaneous or epidural administration of medications.

Reason

Due to an increase in complaints for leaks at the distal male Luer and Luer cap. Baxter implemented enhancements to the blue winged cap to improve the seal between the cap and the Luer.

Code Info

Product Codes: 2C1071KJP, 2C1073KJP, 2C1075KJP, 2C1080KJP, 2C1082KJP, 2C1955KJP, 2C1976KJ. Expiration Dates: 06/30/13, 12/31/13, 06/30/14, 12/31/14

Distribution

Worldwide Distribution - USA (nationwide), Puerto Rico, and Internationally to Brazil, Colombia, Panama, Mexico, Singapore Japan, Australia, New Zealand, Hong Kong, India, Thailand, Taiwan, and Korea.

Quantity

601,230 total