FDA Enforcement Class II Ongoing

Brand Name: I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) with the below descriptions and corresponding article numbers. 1. Anterior Column Plate; Article Numbers: 21218-16, 21219-16. 2. Anterior Column Plate; Article Numbers: 21218-14, 21219-14. 3. Biplanar 2-Column Plate; Article Numbers: 21216-10, 21216-12, 21217-10, 21217-12. 4. Cancellous Screw; Article Numbers: 30591-36, 30591-70, 30591-75, 30591-80, 30591-85, 30591-90. 5. Cancellous Screw, Locking; Article Numbers: 37422-12-N, 37422-14-N, 37422-16-N, 37422-18-N, 37422-20-N, 37422-22-N, 37422-24-N, 37422-26-N, 37422-28-N, 37422-30-N, 37422-32-N, 37422-34-N, 37422-36-N, 37422-38-N, 37422-40-N, 37422-42-N, 37422-44-N, 37422-46-N, 37422-48-N, 37422-50-N, 37422-55-N, 37422-60-N, 37422-65-N, 37422-70-N, 37422-75-N, 37422-80-N, 37422-85-N, 37422-90-N, 37422-95-N, 37422-100-N, 37422-105-N, 37422-110-N, 37422-115-N, 37422-120-N, 37592-18, 37592-20, 37592-22, 37592-24, 37592-26, 37592-28, 37592-30, 37592-32, 37592-34, 37592-36, 37592-38, 37592-40, 37592-42, 37592-46, 37592-48, 37592-50, 37592-54, 37592-56, 37592-58, 37592-60, 37592-65, 37592-70, 37592-75, 37592-80, 37592-85, 37592-90. 6. Cannulated Screw, Cancellous; Article Numbers: 31851-60, 31851-65, 31851-70, 31851-75, 31851-80, 31851-85, 31851-90, 31851-95, 31851-100, 31851-105, 31851-110, 31851-115, 31851-120, 31851-125, 31851-130, 31851-135, 31851-140, 31851-145, 31851-150, 31851-155, 31851-160, 31851-165, 31851-170, 31851-175, 31851-180, 31851-185, 31851-190, 31851-195, 31851-200, 31854-60, 31854-65, 31854-70, 31854-75, 31854-80, 31854-85, 31854-90, 31854-95, 31854-100, 31854-105, 31854-110, 31854-115, 31854-120, 31854-125, 31854-130, 31854-135, 31854-140, 31854-145, 31854-150, 31854-155, 31854-160, 31854-165, 31854-170, 31854-175, 31854-180, 31854-185, 31854-190, 31854-195, 31854-200. 7. Cortical Screw; Article Numbers: 32351-10, 32351-12, 32351-14, 32351-16, 32351-18, 32351-20, 32351-22, 32351-24, 32351-26, 32351-28, 32351-30, 32351-32, 32351-34, 32351-36, 32351-38, 32351-40, 32351-42, 32351-44, 32351-46, 32351-48, 32351-50, 32351-55, 32351-60, 32351-65, 32351-70, 32351-75, 32351-80, 32351-85, 32352-16, 32352-18, 32352-20, 32352-22, 32352-24, 32352-26, 32352-28, 32352-30, 32352-32, 32352-34, 32352-36, 32352-38, 32352-40, 32352-42, 32352-44, 32352-46, 32352-48, 32352-50, 32352-55, 32352-60, 32352-65, 32352-70, 32352-75, 32352-80, 32352-85, 32352-90, 32352-95, 32352-100, 32352-105, 32352-110, 32352-115, 32352-120, 32455-18, 32455-20, 32455-22, 32455-24, 32455-26, 32455-28, 32455-30, 32455-32, 32455-34, 32455-36, 32455-38, 32455-40, 32455-42, 32455-44, 32455-46, 32455-48, 32455-50, 32455-52, 32455-54, 32455-56, 32455-58, 32455-60, 32455-65, 32455-70, 32455-75, 32455-80, 32455-85, 32455-90, 32475-32. 8. Cortical Screw, Locking; Article Numbers: 37455-18, 37455-20, 37455-22, 37455-24, 37455-26, 37455-28, 37455-30, 37455-32, 37455-34, 37455-36, 37455-38, 37455-40, 37455-42, 37455-44, 37455-46, 37455-48, 37455-50, 37455-52, 37455-54, 37455-56, 37455-60, 37455-65, 37455-70, 37455-75, 37455-80, 37455-85, 37455-90. 9. Curved Plate; Article Numbers: 211911-4, 211911-5, 211911-6, 211911-7, 211911-8, 211911-10, 211911-12, 211911-14, 211911-16, 211911-20. 10. Posterior Column Plate, Left; Article Numbers: 212121-6. 11. Posterior Column Plate, Right; Article Numbers: 212131-6. 12. Posterior Wall Plate, Extended, Left; Article Numbers: 212101-SM. 13. Posterior Wall Plate, Extended, Left; Article Numbers: 212101-LA. 14. Posterior Wall Plate, Extended, Right; Article Numbers: 212111-LA, 212111-SM. 15. Posterior Wall Plate, Left; Article Numbers: 212141-7, 212141-8. 16. Posterior Wall Plate, Right; Article Numbers: 212151-7, 212151-8. 17. PRS Plate; Article Numbers: 21194-6, 21194-7, 21194-8, 21195-10, 21195-12, 21195-7. 18. Quadrilateral Buttress Plate; Article Numbers: 21208-13, 21209-13. 19. Quadrilateral Buttress Plate; Article Numbers: 21208-11, 21209-11. 20. Rim Plate, Left; Article

Recall: Z-1608-2026 · Reported April 1, 2026

Enforcement

Recall Number
Z-1608-2026
Event ID
98341
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
I.T.S. GmbH
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
April 1, 2026
Initiation Date
January 23, 2026
Classification Date
March 20, 2026
Address
Autal 28, Lasnitzhohe, N/A, Austria

Description

Brand Name: I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) with the below descriptions and corresponding article numbers. 1. Anterior Column Plate; Article Numbers: 21218-16, 21219-16. 2. Anterior Column Plate; Article Numbers: 21218-14, 21219-14. 3. Biplanar 2-Column Plate; Article Numbers: 21216-10, 21216-12, 21217-10, 21217-12. 4. Cancellous Screw; Article Numbers: 30591-36, 30591-70, 30591-75, 30591-80, 30591-85, 30591-90. 5. Cancellous Screw, Locking; Article Numbers: 37422-12-N, 37422-14-N, 37422-16-N, 37422-18-N, 37422-20-N, 37422-22-N, 37422-24-N, 37422-26-N, 37422-28-N, 37422-30-N, 37422-32-N, 37422-34-N, 37422-36-N, 37422-38-N, 37422-40-N, 37422-42-N, 37422-44-N, 37422-46-N, 37422-48-N, 37422-50-N, 37422-55-N, 37422-60-N, 37422-65-N, 37422-70-N, 37422-75-N, 37422-80-N, 37422-85-N, 37422-90-N, 37422-95-N, 37422-100-N, 37422-105-N, 37422-110-N, 37422-115-N, 37422-120-N, 37592-18, 37592-20, 37592-22, 37592-24, 37592-26, 37592-28, 37592-30, 37592-32, 37592-34, 37592-36, 37592-38, 37592-40, 37592-42, 37592-46, 37592-48, 37592-50, 37592-54, 37592-56, 37592-58, 37592-60, 37592-65, 37592-70, 37592-75, 37592-80, 37592-85, 37592-90. 6. Cannulated Screw, Cancellous; Article Numbers: 31851-60, 31851-65, 31851-70, 31851-75, 31851-80, 31851-85, 31851-90, 31851-95, 31851-100, 31851-105, 31851-110, 31851-115, 31851-120, 31851-125, 31851-130, 31851-135, 31851-140, 31851-145, 31851-150, 31851-155, 31851-160, 31851-165, 31851-170, 31851-175, 31851-180, 31851-185, 31851-190, 31851-195, 31851-200, 31854-60, 31854-65, 31854-70, 31854-75, 31854-80, 31854-85, 31854-90, 31854-95, 31854-100, 31854-105, 31854-110, 31854-115, 31854-120, 31854-125, 31854-130, 31854-135, 31854-140, 31854-145, 31854-150, 31854-155, 31854-160, 31854-165, 31854-170, 31854-175, 31854-180, 31854-185, 31854-190, 31854-195, 31854-200. 7. Cortical Screw; Article Numbers: 32351-10, 32351-12, 32351-14, 32351-16, 32351-18, 32351-20, 32351-22, 32351-24, 32351-26, 32351-28, 32351-30, 32351-32, 32351-34, 32351-36, 32351-38, 32351-40, 32351-42, 32351-44, 32351-46, 32351-48, 32351-50, 32351-55, 32351-60, 32351-65, 32351-70, 32351-75, 32351-80, 32351-85, 32352-16, 32352-18, 32352-20, 32352-22, 32352-24, 32352-26, 32352-28, 32352-30, 32352-32, 32352-34, 32352-36, 32352-38, 32352-40, 32352-42, 32352-44, 32352-46, 32352-48, 32352-50, 32352-55, 32352-60, 32352-65, 32352-70, 32352-75, 32352-80, 32352-85, 32352-90, 32352-95, 32352-100, 32352-105, 32352-110, 32352-115, 32352-120, 32455-18, 32455-20, 32455-22, 32455-24, 32455-26, 32455-28, 32455-30, 32455-32, 32455-34, 32455-36, 32455-38, 32455-40, 32455-42, 32455-44, 32455-46, 32455-48, 32455-50, 32455-52, 32455-54, 32455-56, 32455-58, 32455-60, 32455-65, 32455-70, 32455-75, 32455-80, 32455-85, 32455-90, 32475-32. 8. Cortical Screw, Locking; Article Numbers: 37455-18, 37455-20, 37455-22, 37455-24, 37455-26, 37455-28, 37455-30, 37455-32, 37455-34, 37455-36, 37455-38, 37455-40, 37455-42, 37455-44, 37455-46, 37455-48, 37455-50, 37455-52, 37455-54, 37455-56, 37455-60, 37455-65, 37455-70, 37455-75, 37455-80, 37455-85, 37455-90. 9. Curved Plate; Article Numbers: 211911-4, 211911-5, 211911-6, 211911-7, 211911-8, 211911-10, 211911-12, 211911-14, 211911-16, 211911-20. 10. Posterior Column Plate, Left; Article Numbers: 212121-6. 11. Posterior Column Plate, Right; Article Numbers: 212131-6. 12. Posterior Wall Plate, Extended, Left; Article Numbers: 212101-SM. 13. Posterior Wall Plate, Extended, Left; Article Numbers: 212101-LA. 14. Posterior Wall Plate, Extended, Right; Article Numbers: 212111-LA, 212111-SM. 15. Posterior Wall Plate, Left; Article Numbers: 212141-7, 212141-8. 16. Posterior Wall Plate, Right; Article Numbers: 212151-7, 212151-8. 17. PRS Plate; Article Numbers: 21194-6, 21194-7, 21194-8, 21195-10, 21195-12, 21195-7. 18. Quadrilateral Buttress Plate; Article Numbers: 21208-13, 21209-13. 19. Quadrilateral Buttress Plate; Article Numbers: 21208-11, 21209-11. 20. Rim Plate, Left; Article

Reason

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Code Info

1. Article Number (Lot Numbers): 21218-16 (679/112166, 679/122229, 679/122230, 679/122304), 21219-16 (679/012441, 679/062456, 679/0624561, 679/112167, 679/122228). 2. Article Number (Lot Numbers): 21218-14(679/032417, 679/102231, 679/112158, 679/112164, 679/122231), 21219-14 (679/012505, 679/112159, 679/112165, 679/122227, 679/122305). 3. Article Number (Lot Numbers): 21216-10 (679/032501, 679/062453, 679/072276, 679/072510, 679/092203, 679/092401, 679/112168, 679/112409, 679/122270), 21216-12 (679/032503, 679/062344, 679/062454, 679/072277, 679/092204, 679/112170, 679/112413, 679/122271), 21217-10 (679/032502, 679/062346, 679/072278, 679/092402, 679/092404, 679/102216, 679/112169, 679/112423), 21217-12 (679/022238, 679/062345, 679/072513, 679/092403, 679/112171, 679/122407, 679/122426). 4. Article Number (Lot Numbers): 30591-36 (18/922090), 30591-70 (18/231539, 18/775273), 30591-75 (18/751307), 30591-80 (18/777295), 30591-85 (18/301308), 30591-90 (18/311309). 5. Article Number (Lot Numbers): 37422-12-N (18/145897, 18/461842), 37422-14-N (18/215281, 679/032001, 679/062320, 679/092355, 679/122326), 37422-16-N (679/052229, 679/052307, 679/102214, 679/102306, 679/122329), 37422-18-N (18/240520, 18/452529, 18/473354, 679/052316, 679/102215, 679/102337, 679/122330), 37422-20-N (679/0323147, 679/072306, 679/102211, 679/102338, 679/122331), 37422-22-N (18/092876, 18/543636, 679/012406, 679/062361, 679/072312, 679/102212, 679/102348), 37422-24-N (679/012407, 679/052333, 679/072313, 679/102349), 37422-26-N (679/012435, 679/062316, 679/072322, 679/102353), 37422-28-N (18/192499, 679/012422, 679/032211, 679/062317, 679/062323, 679/072323, 679/102361), 37422-30-N (18/217289, 679/012436, 679/032212, 679/0323142, 679/072324, 679/102354), 37422-32-N (18/218290, 679/012437, 679/052335, 679/052337, 679/072329, 679/112306), 37422-34-N (18/031830, 18/219291, 679/012446, 679/062337, 679/062341, 679/072330, 679/112318), 37422-36-N (18/220292, 18/240540, 18/734742, 679/012447, 679/042317, 679/042321, 679/062003, 679/072334, 679/102361), 37422-38-N (18/105881, 18/221293, 18/362296, 18/412482, 679/012243, 679/022407, 679/052331, 679/052334, 679/072349, 679/112323), 37422-40-N (18/052830, 18/222294, 18/240538, 18/493069, 18/736744, 679/022401, 679/032310, 679/042302, 679/072354, 679/112329), 37422-42-N (18/053831, 18/081544, 18/223295, 18/364298, 18/733745, 679/022402, 679/062371, 679/082304, 679/112330), 37422-44-N (18/054832, 18/224296, 18/365299, 18/737746, 18/781832, 679/022412, 679/042331, 679/042345, 679/082314, 679/112335), 37422-46-N (18/052578, 18/225314, 18/366300, 18/738747, 679/022419, 679/042313, 679/042318, 679/082319, 679/112336), 37422-48-N (18/056834, 18/226315, 679/022423, 679/052324, 679/052328, 679/072316, 679/092314, 679/112337), 37422-50-N (18/227316, 18/740748, 679/022436, 679/042336, 679/052314, 679/092342, 679/112338), 37422-55-N (18/113882, 18/368302, 679/032215, 679/052336, 679/062309, 679/092322, 679/112324), 37422-60-N (18/229318, 18/240517, 18/927095, 28/33034, 679/032403, 679/062349, 679/062351, 679/092325, 679/122307), 37422-65-N (18/118884, 18/230319, 18/271754, 18/743749, 18/761192, 28/35846, 679/032216, 679/072303, 679/092326, 679/122308), 37422-70-N (679/102213, 18/744750, 18/893916, 679/062368, 679/092338, 679/102213, 679/122309), 37422-75-N (18/122888, 18/231839, 679/072304, 679/122306), 37422-80-N (679/122254, 18/124887, 679/062352, 679/122320), 37422-85-N (18/056936, 18/271837, 679/012256, 679/092339), 37422-90-N (18/050600, 18/321273, 679/092340), 37422-95-N (18/130891, 679/092341), 37422-100-N (18/132892, 18/341607, 679/102403), 37422-105-N (679/102404), 37422-110-N (18/136894, 18/371843, 679/092337), 37422-115-N (18/138895, 18/391845, 679/012262), 37422-120-N (18/140896, 679/092349), 37592-18 (679/072351), 37592-20 (18/831792, 18/252862), 37592-22 (18/172773, 679/072356), 37592-24 (679/082301), 37592-26 (18/182774, 679/082303), 37592-28 (679/012140, 679/082310), 37592-30 (18/112775, 679/082317), 37592-32 (679/082312), 37592-34 (18/192863, 679/082313), 37592-36 (679/082318), 37592-38 (679/092308), 37592-40 (679/052116, 679/072314), 37592-42 (679/092310), 37592-46 (18/092772, 679/092312), 37592-48 (679/052118, 679/092313), 37592-50 (679/092343), 37592-54 (679/072315), 37592-56 (18/721587), 37592-58 (679/092334), 37592-60 (679/042213, 679/072325, 679/092345), 37592-65 (679/072331), 37592-70 (679/042289, 679/072340, 679/092353), 37592-75 (18/429442, 679/092351), 37592-80 (679/052204, 679/072341, 679/092335), 37592-85 (679/042291, 679/072342, 679/092354), 37592-90 (679/042292, 679/092352). 6. Article Number (Lot Numbers): 31851-60 (18/373463, 18/813133), 31851-65 (18/374464), 31851-70 (18/375465), 31851-75 (18/376467), 31851-80 (18/125212, 18/377468, 18/431563, 18/867967), 31851-85 (18/044945, 18/378469, 18/441564), 31851-90 (18/045946, 18/343266, 18/451565), 31851-95 (18/046947, 18/128224, 18/379470, 18/461566), 31851-100 (18/129225, 18/471567, 18/821469, 18/860968), 31851-105 (18/380471), 31851-110 (18/901298, 18/387472), 31851-115 (18/381473, 18/048949), 31851-120 (18/049950, 18/133227, 18/511577, 18/862970), 31851-125 (18/050951, 18/347474, 18/521578), 31851-130 (18/051952, 18/344368, 18/531579), 31851-135 (18/382475, 18/541580), 31851-140 (18/052953, 18/551777, 18/863971), 31851-145 (18/054954, 18/137229, 18/345351, 18/561779), 31851-150 (18/138260, 18/191042, 18/346352, 18/571780, 18/797898), 31851-155 (18/201043, 18/384476), 31851-160 (18/062955, 18/799900, 18/866972), 31851-165 (18/063956, 18/211045, 18/385477), 31851-170 (18/141230, 18/221046, 18/348353), 31851-175 (18/064957, 18/142263, 18/386398), 31851-180 (18/065958, 18/143264, 18/388399), 31851-185 (18/389400), 31851-190 (18/066959, 18/390401), 31851-195 (18/067960, 18/391478), 31851-200 (18/013266, 18/392479), 31854-60 (18/393482), 31854-65 (18/394499, 18/691605), 31854-70 (18/070983, 18/395500), 31854-75 (18/071984, 18/396501), 31854-80 (18/072985, 18/397502, 18/731608, 18/871960), 31854-85 (18/149247, 18/240529, 18/741609, 18/872961), 31854-90 (18/150280, 18/398503, 18/751610, 18/873962), 31854-95 (18/399504, 18/761611, 18/874963), 31854-100 (18/400505, 18/771612, 18/868964), 31854-105 (18/401506), 31854-110 (18/073986, 18/869965), 31854-115 (18/074987, 18/402507, 18/801615), 31854-120 (18/403508, 18/870966), 31854-125 (18/404509, 18/821617), 31854-130 (18/075988, 18/405366, 18/831618), 31854-135 (18/076989, 18/406510), 31854-140 (18/407369, 18/851693), 31854-145 (18/408370, 18/861706), 31854-150 (18/160285, 18/409371, 18/871704), 31854-155 (18/410511), 31854-160 (18/1622951, 18/411372, 18/862761, 18/891708), 31854-165 (18/163286, 18/412512), 31854-170 (18/164296, 18/413513), 31854-175 (18/079992, 18/414514), 31854-180 (18/080993, 18/415515), 31854-185 (18/081994, 18/416373, 18/808929), 31854-190 (18/082995), 31854-195 (18/083996, 18/961715), 31854-200 (18/084997, 18/417516). 7. Article Number (Lot Numbers): 32351-10 (18/019937, 18/208200, 18/731287, 679/052205, 679/122249), 32351-12 (18/020938, 18/209211, 18/741316, 679/062203, 679/102243), 32351-14 (18/210212, 18/501230, 18/761312, 679/012324, 679/032524, 679/042419, 679/062225, 679/082501, 679/122207), 32351-16 (18/022940, 18/211212, 18/771313, 679/022311, 679/022533, 679/032525, 679/042421, 679/072201, 679/092501, 679/122204), 32351-18 (18/212213, 18/781322, 18/801041, 679/022312, 679/032526, 679/042427, 679/072202, 679/102501, 679/122208), 32351-20 (18/024942, 18/181975, 18/213214, 18/791323, 679/032527, 679/042428, 679/102502, 679/112225), 32351-22 (18/025943, 18/214215, 18/801324, 18/8013241, 18/821061, 679/052219, 679/112235), 32351-24 (18/026988, 18/215216, 18/811325, 18/8113251, 18/831062, 679/052206, 679/102239), 32351-26 (18/027989, 18/216217, 18/821326, 18/841063, 679/052202, 679/052407, 679/112236), 32351-28 (18/028990, 18/217218, 18/8313271, 679/052220, 679/102235), 32351-30 (18/029991, 18/218219, 18/841314, 679/062205, 679/122260), 32351-32 (18/219220, 18/301035, 18/851315, 18/8513151, 679/012303, 679/052203), 32351-34 (18/220221, 18/311041, 18/861328, 18/961089, 679/052243, 679/112229), 32351-36 (18/230222, 18/411090, 679/052221, 679/112241), 32351-38 (18/131369, 18/1313691, 18/221226, 18/321091, 26/F66144, 679/052207, 679/102247), 32351-40 (18/222223, 18/331092, 679/052208, 679/122243), 32351-42 (18/223224, 18/341093, 18/871329, 18/8713291, 679/112214), 32351-44 (18/351094, 18/881330, 18/8813301, 18/921138, 679/052209, 679/112211), 32351-46 (18/225227, 18/361095, 18/891331, 18/8913311, 18/931313, 679/042217, 679/112215), 32351-48 (18/226228, 18/371096, 679/052210, 679/122244), 32351-50 (18/227229, 18/381097, 18/941314, 679/012304, 679/052211), 32351-55 (18/391098, 18/901332, 18/951142, 18/951315, 679/042218, 679/112237), 32351-60 (18/229231, 18/401099, 18/961316, 679/042283, 679/122250), 32351-65 (18/981144, 679/022326, 679/042284), 32351-70 (18/236233), 32351-75 (18/237234, 679/022328), 32351-80 (18/461103, 679/042286), 32351-85 (18/431104), 32352-16 (679/022406, 679/072207, 679/122427), 32352-18 (679/012501, 679/022344, 679/052182), 32352-20 (18/670753, 679/012453, 679/012504, 679/022345, 679/052184, 679/052311), 32352-22 (18/237340, 679/012454, 679/022346, 679/022506, 679/062101, 679/072208, 679/122003), 32352-24 (18/238437, 18/672755, 679/022404, 679/032379, 679/032528, 679/052312, 679/062102, 679/072209, 679/122424), 32352-26 (18/239357, 18/673756, 679/022352, 679/022405, 679/032512, 679/052313, 679/072210, 679/122416), 32352-28 (18/240358, 18/674757, 679/022411, 679/032267, 679/032324, 679/032529, 679/052303, 679/072211, 679/122418), 32352-30 (18/675758, 18/950241, 679/022417, 679/022513, 679/032325, 679/052304, 679/072213, 679/102414, 679/102513), 32352-32 (18/242360, 18/676759, 679/022418, 679/032323, 679/032530, 679/052305, 679/072214, 679/122421), 32352-34 (18/243361, 18/677760, 679/022426, 679/022514, 679/032326, 679/052301, 679/072280, 679/102401), 32352-36 (18/244362, 18/678761, 679/032327, 679/032531, 679/042274, 679/052302, 679/072282, 679/122422), 32352-38 (18/245363, 18/679762, 679/022428, 679/0323112, 679/032392, 679/032513, 679/072283, 679/112502, 679/122413), 32352-40 (18/246364, 18/680763, 679/012506, 679/022438, 679/032268, 679/032375, 679/042342, 679/072112, 679/072285, 679/112505, 679/122419), 32352-42 (18/247365, 18/681764, 679/012507, 679/012508, 679/032347, 679/042343, 679/062246), 32352-44 (18/248366, 18/682765, 679/012217, 679/012503, 679/032339, 679/062247), 32352-46 (18/249367, 18/683766, 18/819818, 679/012218, 679/022526, 679/032371, 679/042332, 679/062249, 679/072294, 679/102402, 679/122501), 32352-48 (18/250368, 679/022444, 679/022505, 679/032372, 679/062251, 679/102180), 32352-50 (18/251369, 18/685768, 679/012509, 679/022452, 679/032364, 679/042329, 679/062253, 679/072295, 679/112434), 32352-55 (18/252370, 18/686769, 679/022504, 679/032376, 679/062255, 679/072296, 679/072297, 679/122401, 679/122504), 32352-60 (18/253438, 18/687770, 679/012222, 679/022527, 679/032328, 679/062257, 679/072298), 32352-65 (18/254460, 18/688798, 679/022457, 679/022520, 679/032343, 679/062259), 32352-70 (18/255371, 18/823821, 679/032344, 679/032408, 679/042316, 679/062261, 679/082201), 32352-75 (18/256429, 679/022522, 679/032398, 679/032399, 679/032409, 679/062263), 32352-80 (18/257430, 679/022523, 679/0323104, 679/0323107, 679/032410, 679/062265), 32352-85 (679/0323109, 679/0323113, 679/062267), 32352-90 (679/0323138, 679/0323141, 679/062268), 32352-95 (679/0323146, 679/0323148, 679/062271), 32352-100 (679/012230, 679/032266, 679/032411, 679/042301, 679/062273), 32352-105 (679/032412, 679/032509, 679/042311, 679/062275), 32352-110 (679/032413, 679/032521, 679/042312, 679/062277), 32352-115 (679/032418, 679/042319, 679/062279), 32352-120 (679/032419, 679/042320, 679/062281), 32455-18 (18/371232), 32455-20 (18/892124), 32455-22 (18/391234, 18/434375, 679/032395), 32455-24 (18/401235, 18/435376, 26/F62579, 679/032387, 679/032396, 679/092502), 32455-26 (18/411236, 18/436377, 679/032397, 679/092503), 32455-28 (18/437378, 679/032345), 32455-30 (18/256241, 679/032346, 679/032388), 32455-32 (18/438380, 18/441239, 26/F62581, 679/032384), 32455-34 (18/440381, 18/451240, 18/565613, 26/F62918, 679/032385), 32455-36 (18/441382, 18/461241, 26/F62582, 679/032389), 32455-38 (18/442383, 18/471242, 679/022353, 679/032363, 679/092507), 32455-40 (18/481243, 18/4812431, 26/F62583, 26F/62583, 679/022351, 679/032349), 32455-42 (18/491244, 26/F62584), 32455-44 (18/445386, 18/501245, 679/022354, 679/032366), 32455-46 (18/511246, 26/F62585, 679/032317, 679/032367), 32455-48 (18/231444, 18/446387, 679/022355, 679/032368), 32455-50 (18/531248), 32455-52 (18/541249, 679/032318, 679/032369), 32455-54 (18/377660, 18/392952, 18/551250), 32455-56 (679/032373), 32455-58 (18/378662, 18/412953, 18/571252), 32455-60 (679/032319, 679/032374), 32455-65 (18/591254), 32455-70 (18/601255, 26/F62586), 32455-75 (18/611256), 32455-80 (18/621257, 26/F62587), 32455-85 (18/231516, 26/F62588), 32455-90 (18/641259, 679/0323108), 32475-32 (26/F16726). 8. Article Number (Lot Numbers): 37455-18 (18/231558), 37455-20 (18/021319), 37455-22 (18/211320), 37455-24 (18/671425, 18/231558), 37455-26 (18/221322, 18/231449), 37455-28 (18/041323), 37455-30 (18/231324), 37455-32 (18/051325), 37455-34 (18/241326), 37455-36 (18/061375, 26/F62605), 37455-38 (18/251376), 37455-40 (18/532513, 18/691442, 26/F62606), 37455-42 (18/151343, 18/261380), 37455-44 (18/081388), 37455-46 (18/231515), 37455-48 (18/231514), 37455-50 (18/281394), 37455-52 (18/101395), 37455-54 (18/701444), 37455-56 (18/111408, 18/656368), 37455-60 (18/121449, 18/572515), 37455-65 (18/131479), 37455-70 (18/141480, 26/F62607), 37455-75 (18/151481, 26/F62608), 37455-80 (18/686476, 26/F62609), 37455-85 (18/171483, 26/F62610), 37455-90 (18/181484). 9. Article Number (Lot Numbers): 211911-4 (00/0825-10, 679/012425, 679/022531, 679/0323131, 679/062430, 679/0624301, 679/102431, 679/122267), 211911-5 (00/0724-28, 00/1225-04, 679/012426, 679/022321, 679/022518, 679/0323132, 679/052527, 679/0624311, 679/0624312, 679/102432), 211911-6 (00/1024-15, 679/012427, 679/022322, 679/022501, 679/042513, 679/062432, 679/0624322, 679/102433, 679/112177), 211911-7 (00/0724-31, 00/0724-32, 00/0724-33, 00/0724-34, 00/0724-35, 00/0725-16, 00/1225-05, 679/022323, 679/022502, 679/052529, 679/062231, 679/102434), 211911-8 (679/012428, 679/022503, 679/022532, 679/0323135, 679/112178, 679/112407), 211911-10 (00/0724-39, 00/1024-17, 679/0323127, 679/102429, 679/112179, 679/122266), 211911-12 (679/0323128, 679/102228, 679/112203, 679/112219), 211911-14 (679/0323129, 679/102229, 679/112221), 211911-16 (679/0323130, 679/062232, 679/112222), 211911-20 (679/062429, 679/112418, 679/122268). 10. Article Number (Lot Numbers): 212121-6 (679/052152, 679/062338, 679/122235). 11. Article Number (Lot Numbers): 212131-6 (679/062319, 679/122234). 12. Article Number (Lot Numbers): 212101-SM (679/042515, 679/052322, 679/122215, 679/122303, 679/122402). 13. Article Number (Lot Numbers): 212101-LA (00/0724-48, 00/1024-09, 679/062313, 679/062443, 679/112206, 679/122302, 679/122405). 14. Article Number (Lot Numbers): 212111-LA (679/032421, 679/042516, 679/062347, 679/062402, 679/062445, 679/122414), 212111-SM (00/0724-65, 00/1024-10, 00/1024-11, 679/022530, 679/032246, 679/062318, 679/062403, 679/062473, 679/092237, 679/112015, 679/122406). 15. Article Number (Lot Numbers): 212141-7 (00/0724-81, 00/0724-85, 679/032331, 679/052168, 679/062448, 679/122214), 212141-8 (679/022408, 679/032332, 679/052429, 679/062449, 679/0624491, 679/062474, 679/122233). 16. Article Number (Lot Numbers): 212151-7 (00/0624-06, 679/012449, 679/052153, 679/062219, 679/062450, 679/062475, 679/102230, 679/112239), 212151-8 (00/0624-11, 00/0724-51, 679/012450, 679/032248, 679/062451, 679/072508). 17. Article Number (Lot Numbers): 21194-6 (679/042247), 21194-7 (679/042249), 21194-8 (679/042250), 21195-10 (679/042252), 21195-12 (679/042253), 21195-7 (679/032221). 18. Article Number (Lot Numbers): 21208-13 (679/032414, 679/052428, 679/062122, 679/102435, 679/112162, 679/112174, 679/122221, 679/122222), 21209-13 (679/062442, 679/0624421, 679/112163, 679/112175, 679/112408, 679/122236). 19. Article Number (Lot Numbers): 21208-11 (679/012429, 679/062440, 679/0624401, 679/112160, 679/112172, 679/122219), 21209-11 (679/062441, 679/112161, 679/112173, 679/112412, 679/122237, 679/122238, 679/122239). 20. Article Number (Lot Numbers): 212061-10 (00/0724-110, 679/062311, 679/112217, 679/122218), 212061-14 (00/0724-42, 00/0724-47, 679/032238, 679/032337, 679/062312). 21. Article Number (Lot Numbers): 212071-10 (679/032415, 679/052329), 212071-14 (679/032338, 679/032416). 22. Article Number (Lot Numbers): 211751-5 (679/052320, 679/112180, 679/112202, 679/112220), 211771-4 (679/052321, 679/102227, 679/122301). 23. Article Number (Lot Numbers): 211821-10 (679/012423, 679/032333, 679/062426, 679/112176), 211821-11 (00/0724-71, 00/0724-74, 679/032334), 211821-12 (679/022320), 211821-14 (679/012424, 679/122217). 24. Article Number (Lot Numbers): 21161-4 (679/092320, 679/112201), 21161-6 (679/032305, 679/032306, 679/042304, 679/052506, 679/062218), 211621-4 (00/0724-18, 00/0724-63, 679/042305, 679/062423, 679/102226, 679/112414), 211621-6 (00/0825-13, 679/022529, 679/042306, 679/062424, 679/112415). 25. Article Number (Lot Numbers): 211631-6 (00/0125-35, 679/022515, 679/032315, 679/062302, 679/062425, 679/0624251, 679/112411), 211631-8 (00/0724-118, 00/1220-11, 679/062303, 679/122216).

Distribution

US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Quantity

46,637 units