FDA Enforcement Class II Ongoing

GETINGE USA SALES LLC OCEAN Chest Drain System Atrium 2100 mL Model Number: 2002-000

Recall: Z-1607-2022 · Reported August 31, 2022

Enforcement

Recall Number
Z-1607-2022
Event ID
90336
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Mckesson Medical-Surgical Inc. Corporate Office
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
August 31, 2022
Initiation Date
May 25, 2022
Classification Date
August 25, 2022
Address
9954 Maryland Drive, Deep Run Iii Ste. 4000, Richmond, VA, 23233, United States

Description

GETINGE USA SALES LLC OCEAN Chest Drain System Atrium 2100 mL Model Number: 2002-000

Reason

Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

Code Info

GTIN: 00650862100013; 20650862100017

Distribution

US Nationwide

Quantity

6 each