FDA Enforcement Class II Terminated

Nuclear gamma cameras (ADAC VERTEX PLUS/SOLUS/CARDIO, ADAC VERTEX V60 & ADAC CARDIO C60) Nuclear gamma cameras intended to produce images depicting the anatomical distribution of single photon and positron emitting radioisotopes with the human body for interpretation by medical personnel.

Recall: Z-1607-2015 · Reported May 20, 2015

Enforcement

Recall Number
Z-1607-2015
Event ID
71135
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 20, 2015
Initiation Date
March 11, 2015
Classification Date
May 13, 2015
Termination Date
June 26, 2018
Address
595 Miner Rd, N/A, Cleveland, OH, 44143-2131, United States

Description

Nuclear gamma cameras (ADAC VERTEX PLUS/SOLUS/CARDIO, ADAC VERTEX V60 & ADAC CARDIO C60) Nuclear gamma cameras intended to produce images depicting the anatomical distribution of single photon and positron emitting radioisotopes with the human body for interpretation by medical personnel.

Reason

The firm was notified by a customer that the collimator exchange carriage and the collimator storage cabinet were misaligned.

Code Info

VERTEX PLUS/SOLUS/CARDIO Model #88297; Serial #s: 9504103, 9703171, 9802304, 9809124, 9809415, 9810438, 9910015, 9911711, _0002032, 0007001M, 0008009B, 0011002B, 0101002C, 0102019V, 0103020V, 01069VM, 01070VM, 0111001V, 01295VM, 01296K, 01864VM, 03026VM, 0939VM, 100098-01N, 1006, 1008, 10252, 1027, 1062701, 1062701, 10630, 1092, 1097, 11, 11090, 130234-01NVERTEX, 2035, 2038, 2047, 3059, 3060, 3064, 4007, 4078, 5934V, 6052, 6090VM, 6261, 7373VM, 7376VM, 7386VM, 7560VM, 7561VM, 8002, 802080322, 85247, 9054VM/V, 9056VM/V, 9073152, 9130VM, 9205345, 940384, 9408127, 9408136, 9503084, 9504096, 9506016, 9506145, 9511123, 9512008, 9512009, 9512018, 9601012, 9601015, 9601016, 9601024, 9601036, 9602161, 9602245, 9602246, 9602254, 9603336, 9604016, 9604021, 9605024, 9605026, 9605037, 9606211, 9607015, 9607024, 9608031, 9608046, 9608049, 9608050, 9608052, 9608052, 9608064, 9609063, 9609081, 9610058, 9611003, 9611035, 9611077, 9611082, 9612072, 9701127, 9702095, 9702150, 9703154, 9703165, 9704044, 9704178, 9704183, 9705004, 9705186, 9706005, 9706182, 9706189, 9707002, 9707047, 9707050, 9707054, 9708062, 9708065, 9708069, 9708076, 9708078, 9708084, 9708088, 9708227, 9709022, 9709061, 9710121, 9710127, 9710244, 9711025, 9711034, 9711036, 9711074, 9711078, 9711261, 9711271, 9801047, 9801227, 9801232, 9801297, 9801299-CN, 9802050, 9802255, 9802260, 9802300, 9802306, 9802397, 9803003, 9803014, 9803067, 9803083, 9803269, 9803272, 9803288, 9803290, 9803293, 9803295, 9803303, 9803312, 9803316, 9803322, 9804069, 9804330, 9804333, 9804342, 9805344, 9805350, 9805376, 9806006, 9806014, 9806363, 9806373A, 9806377, 9806388, 98066002, 9807002, 9807391, 9807421, 9808036, 9808090, 9808108, 9808451, 9808452, 9808457, 9808480, 9808829, 9809065-CN, 9809067, 9809113, 9809122, 9809125, 9809498, 9809687, 9810075, 9810136, 9810428ADAC#7973VM, 9810442, 9810551, 9810557, 9811020, 9811093, 9811569, 9811737, 9812582, 9901162, 9901597, 9901605, 9902150, 9902622, 9902630, 9903632, 9903637, 9903641, 9904089, 9905003, 9905012, 9905659, 9905667, 9906670, 9907016, 9907017, 9907019, 9907029, 9909011, 9909161, 9910017, 9910031, 9911458, 9911698, 9911702, 9911711, 9912181, 9912463, AD1261, AMT00848VM, B0006129, B0007273, B0008001, B0008003, B0008012, B0008021, B0008030, B001200212, B0102114, B0104125, B0104128, B0105138, B0106169, B01070191, B01070192, B0107179, B01080199, B01080204, B01080207, B01090227, B01090228, B01100231, B01100232, B01100233, B01100236, B01100237, B01100238, B01100241, B01100242, B01110251, B01110252, B02050071, B02050262, B02050263, B02050265, B02060167, B02060270, B02060279, B0206027S, B02070289, B02070290, B02070291, B02070297, B02070300, B0207296, B02080318, B0208312, B02090327, B02090333, B02090334, B02100345, B02110348, B02120352, B03010355, B03010359, B03020048, B03080380, C0006050, C0006051, C0007060, C0008070, C0009075, C0101099, C0102103, C0102108, C0104118, C0107145, C01080152, C01080153, C01080154, C02100218, C02110220, C02120223, C03010227, C03050230, C03060193, CANCEL_9809065, CNF, M@10458, M@10482, M@10547, M@105561, M@10573, M@10582, M@10589, M@10591, M@10676, M@10772, M@10837, M@10839, M@10848, M@10850, M@10871, M@10878, M@10882, M@10927, M@10937, M@10967, M@10979, M@11015, M@11020, M@11036, M@11103, M@11103, M@11109, M@11140, M@11178, M@11184, M@11381, M@11398, M@11423, M@11498, M@11525, M@11571, M@11590, M@11609, M@11644, M@11881, M@11939, M@11949, M@11957, M@11972, M@12060, M@12219, M@12412, M@12591, M@12711, M@12787, M@13010, M@13296, M@13466, M@13498, M@13543, M@13549, M@13663, M@13665, M@13688, M@13691, M@13748, M@13813, M@14062, M@22752, M@28051, M@4750019856, M@5828J, M@5906N, M@5919J, M@5936N, M@5956J, M@6002J, M@6012J, M@6075VM, M@6076K, M@6077K, M@6108VA, M@6123VA, M@6143VA, M@6160VM, M@6183VA, M@6244VA, M@6258J, M@6323VM, M@6324J, M@6330VA, M@6341VA, M@6344VM, M@6345, M@6345VA, M@6346VM, M@6348VM, M@6349VM, M@6365VM, M@6366K, M@6368J, M@6370VA, M@6371VM, M@6375J, M@6376J, M@6460VM, M@6470VA, M@6481VA, M@6496VM, M@6528VA, M@6541VA, M@6542VM, M@6566VM, M@6573VM, M@6596VM, M@6600VA, M@6638VM, M@6639J, M@6675VA, M@6716VM, M@6717VM, M@6762VM, M@6769K, M@6771VM, M@6773VM, M@6776VM, M@6797K, M@6805VM, M@6823VM, M@6840K, M@6841VM, M@6867VM, M@6877VM, M@6886VM, M@6891VA, M@6912K, M@6924VM, M@6931VM, M@6932VM, M@6943VM, M@6946VM, M@6947J, M@6985J, M@7003VA, M@7004VA, M@7005VA, M@7017VM, M@7024VM, M@7034K, M@7042VM, M@7053VM, M@7070VM, M@7073VM, M@7085K, M@7086J, M@7095VM, M@7096VM, M@7099K, M@7100VM, M@7101VM, M@7102VM, M@7108VM, M@7109VM, M@7151VM, M@7166K, M@7171VM, M@7172K, M@7181J, M@7190VM, M@7198VA, M@7205K, M@7215K, M@7216VM, M@7217VM, M@7221VA, M@7222VA, M@7229VM, M@7233VM, M@7237VM, M@7245J, M@7246VM, M@7247VM, M@7250VM, M@7253K, M@7300J, M@7308VM, M@7358VA, M@7376VM, M@7380VM, M@7384VM, M@7390VM, M@7393VM, M@7394VM, M@7410VM, M@7420VM, M@7463VM, M@7464VA, M@7470VM, M@7490VM, M@7494K, M@7494K, M@7501VM, M@7503VM-FU1, M@7504VM, M@7505VM1, M@7514VM, M@7515VM, M@7536VM, M@7537VM, M@7538VM, M@7556VM, M@7557VM, M@7569VM, M@7572VM, M@7591VM, M@7592VM, M@7603K, M@7605VM, M@7606VM, M@7623K, M@7624K, M@7626K, M@7627K, M@7628VA, M@7638VM, M@7643VM, M@7654VM, M@7664VM, M@7665K, M@7673VM, M@7674K, M@7679VM, M@7682K, M@7685VM, M@7691VM, M@7695VM, M@7697VM, M@7723VM, M@7729VM, M@7733VM, M@7739K, M@7740VM, M@7760VM, M@7760VM, M@7761VM, M@7761VM, M@7770VM, M@7776VM, M@7789K, M@7801VM, M@7810VM (9809056), M@7811VM, M@7812VM, M@7819VA, M@7823K, M@7824VM, M@7830K, M@7855J, M@7857VM, M@7860VM, M@7861VM, M@7874VM, M@7877VM, M@7877VM, M@7918K, M@7924K, M@7925K, M@7929VM, M@7930VM, M@7932VM, M@7946J, M@7957VM, M@7961VM, M@7962VM, M@7963K, M@7965VM, M@7968VM, M@7975K, M@7987VM, M@7995VM, M@7997VM, M@8004VM, M@8018J, M@8021VM, M@8044VM, M@8064VM, M@8076K, M@8089VM, M@8096VM, M@8105VM, M@8124VA, M@8124VA, M@8146VM, M@8155VM, M@8178VM, M@8203VM, M@8219VA, M@8220K, M@8240K, M@8241VM, M@8247K, M@8248VM, M@8252VM, M@8284K, M@8301K, M@8302VM, M@8306K, M@8337VM, M@8356K, M@8380VM, M@8394K, M@8399VM, M@8436K, M@8462VM, M@8484VM, M@8491K, M@8510VM, M@8528VM, M@8537VM, M@8540K, M@8544K, M@8551VM, M@8561K, M@8606K, M@8609VM, M@8651VM, M@8683VM, M@8686K, M@8693VM, M@8726VM, M@8738K, M@8759K, M@8771VM, M@8781VM, M@8797K, M@8799RV, M@8800VM, M@8803K, M@8812VM, M@8814VM, M@8817RK, M@8876K, M@8882K, M@8884K, M@8888K, M@8897K, M@8903K, M@8906RK, M@8920VM, M@8922VM, M@8935K, M@8936K, M@8942K, M@8943K, M@8948RV, M@8955K, M@8966K, M@8967K, M@8990K, M@8996VM, M@9000K, M@9032K, M@9041K, M@9046K, M@9049VM, M@9063K, M@9080K, M@9081K, M@9085K, M@9091VM, M@9107RV, M@9118K, M@9129K, M@9131K, M@9140K, M@9166VM, M@9185K, M@9186K, M@9191K, M@9194VM, M@9221VM, M@9222VM, M@9231K, M@9247RV, M@9252VM, M@9267VM, M@9283K, M@9293VM, M@9308K, M@9309RV, M@9316VM, M@9327K, M@9337K, M@9338VM, M@9340VM, M@9344K, M@9349K, M@9360VM, M@9369VM, M@9418VM, M@9426K, M@9434K, M@9442K, M@9452K, M@9471RV, M@9476K, M@9479K, M@9480K, M@9502K, M@9515VM, M@9522K, M@9525K, M@9532VM, M@9542VM, M@9543K, M@9562K, M@9563K, M@9566K, M@9583VM, M@9590K, M@9594K, M@9598K, M@9607VM, M@9609K, M@9622K, M@9639K, M@9646K, M@9653VM, M@9672VM, M@9712K, M@9724VM, M@9730K, M@9765K, M@9769K, M@9799K, M@9816K, M@9831VM, M@9834K, M@9837K, M@AMT1258, M@MCD6416VM, M@MCD7036VM, M@O1317, M@R372820, M@S5195VA, M@S5365-V, M@S5820VM, M@S5907V11, M@S6155VA, M@S7115VM, M@S7156VM, M@S7223K, M@S7559VA, M@S7571VM, M@S7817K, M@S7856K, M@S7864K, M@S7904VM, M@S7941VA, M@S8687K, M0101023, R03010363, S0011003, S610075, S6235VM, STOR_QUO_47521360, UNK-6597VM, UNK-7590VM, UNK-8097J1, UNK-8533VM1, UNK-8860VM, UNK-9868K, UNK-9897K, V0007070, V0010090, V0012099, V0104109, V0104110, V0105112, V0106121, V0106123, V0106124, V01070133, V01090146, V01100151, V01100156, V02050166, V02070174, V02070179, V02080190, V02080194, V02080196, V02080199, V02080200, V02090206, V03060224, V03070229 & V03091241. VERTEX V60 Model #882030; Serial #s: 10588, VETEJ, V02060169, 9707207, 00592V, 00645VA, 9910016, Z00645VA, 95, 9602247, 9803014, PA, 9706005, 00634V, 00970VA, 01697VM, 0371V, 00559VM, 01894VM, Z01798VM, 00657VA, 0, 01190VM, M0007001-CN, 9710247, 9902027, 12001, X0101006, 9707206, 9806019, X0005002, X0010004, 9712009, 9712008, 9607063, 9712001, 9909039, 9803003, X03030012, 9904005, 9602253, 9606038, 9603013, 9603008, V02050163, 9810434, 9604041, V02050165, V0106122, 9605033, 9604028, V0009076, 9704175, 3058, 3061, M0007018, 9807089, 9704177, X0101007, 9808021, V01090142, V01090141, 1189-17, 1273-06, 1274-06, V0005062, 2052, V0010085, V04060250, 1043, V03080235, V03080236, V2070177, 9601004, 0003058, 0003061, V02050163, 9810434, 9604041, 9711277, V02050165, V0106122, 9605033, 9706192, V0009076, 9704175, 9707207, VETEJ, 95, 9803014, 9602247, 0371V, 01697VM, 00645VA, 00559VM, 00634V, Z00645VA, 9610256, 9806379, 9811075, 9312049 & V02060169. CARDIO C60 Model #882040; Serial #s: 01609K, PA, 3035, 9810557, C0006050, C01080154, 9708088, 9685K, M@9080K, M@11103, 88297, 0101002C, M@13663, 9803082, C01070146, C0104118, C03010227, M@9488K & 9808108.

Distribution

Worldwide Distribution - US (nationwide) to the states of: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY., and to the countries of : Australia, Austria, Belgium, Brazil, Canada, China, Costa Rica, Egypt, France, Germany, Hong Kong, Ireland, Italy, Japan, Korea, Lebanon, Netherlands, Norway, Pakistan, Portugal, Russia, Singapore, South Korea, Spain,Thailand and United Kingdom.

Quantity

975 units