FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Cancellous Stabilization Screw; Article Numbers: 37303-10, 37303-12, 37303-14, 37303-16, 37303-18, 37303-20, 37303-22, 37303-24, 37303-26, 37303-28, 37303-30. 2. FLS - Foot Locking Plates System, Cortical Screw; Article Numbers: 32271-10, 32271-12, 32271-14, 32271-16, 32271-18, 32271-20, 32271-22, 32271-24, 32271-26, 32271-28, 32271-32, 32271-34. 3. FLS - Foot Locking Plates System, Cortical Stabilization Screw; Article Numbers: 37304-10, 37304-12, 37304-14, 37304-16, 37304-18, 37304-20, 37304-22, 37304-24. 4. FLS - Foot Locking Plates System, FLS Plate, Curved; Article Numbers: 21413-6. 5. FLS - Foot Locking Plates System, FLS Plate, H-Shape; Article Numbers: 21431-ME. 6. FLS - Foot Locking Plates System, FLS Plate, L-Shape; Article Numbers: 21482, 21483. 7. FLS - Foot Locking Plates System, FLS Plate, Straight; Article Numbers: 21421-5. 8. FLS - Foot Locking Plates System, Stabilization Screw; Article Numbers: 37241-10, 37241-12, 37241-14, 37241-16, 37241-18, 37241-20, 37241-22, 37241-24, 37241-26, 37241-28, 37241-30. 9. HOL - Hallux Osteotomy Locking Plate, Hallux Osteotomy Plate; Article Numbers: 21015-08, 21015-09.
Enforcement
- Recall Number
- Z-1605-2026
- Event ID
- 98341
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- I.T.S. GmbH
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 1, 2026
- Initiation Date
- January 23, 2026
- Classification Date
- March 20, 2026
- Address
- Autal 28, Lasnitzhohe, N/A, Austria
Description
FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Cancellous Stabilization Screw; Article Numbers: 37303-10, 37303-12, 37303-14, 37303-16, 37303-18, 37303-20, 37303-22, 37303-24, 37303-26, 37303-28, 37303-30. 2. FLS - Foot Locking Plates System, Cortical Screw; Article Numbers: 32271-10, 32271-12, 32271-14, 32271-16, 32271-18, 32271-20, 32271-22, 32271-24, 32271-26, 32271-28, 32271-32, 32271-34. 3. FLS - Foot Locking Plates System, Cortical Stabilization Screw; Article Numbers: 37304-10, 37304-12, 37304-14, 37304-16, 37304-18, 37304-20, 37304-22, 37304-24. 4. FLS - Foot Locking Plates System, FLS Plate, Curved; Article Numbers: 21413-6. 5. FLS - Foot Locking Plates System, FLS Plate, H-Shape; Article Numbers: 21431-ME. 6. FLS - Foot Locking Plates System, FLS Plate, L-Shape; Article Numbers: 21482, 21483. 7. FLS - Foot Locking Plates System, FLS Plate, Straight; Article Numbers: 21421-5. 8. FLS - Foot Locking Plates System, Stabilization Screw; Article Numbers: 37241-10, 37241-12, 37241-14, 37241-16, 37241-18, 37241-20, 37241-22, 37241-24, 37241-26, 37241-28, 37241-30. 9. HOL - Hallux Osteotomy Locking Plate, Hallux Osteotomy Plate; Article Numbers: 21015-08, 21015-09.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
1. Article Number (Lot Numbers): 37303-10 (18/151720, 18/268729, 679/032432, 679/072268, 679/102237, 679/102307), 37303-12 (18/570637, 679/032434, 679/072269, 679/102346, 679/112227), 37303-14 (18/751091, 26/F66913, 679/012510, 679/062288, 679/102359, 679/112410), 37303-16 (679/072263, 679/102245, 18/151841, 26/F66914, 18/331193, 18/543625, 679/012511, 679/042403, 679/102360), 37303-18 (18/151942, 18/210467, 18/231415, 18/341194, 18/572633, 18/771093, 679/012512, 679/022507, 679/042404, 679/102248, 679/112319), 37303-20 (18/152043, 18/211468, 18/351195, 18/573634, 18/701585, 18/781094, 679/022508, 679/042420, 679/112216, 679/112328), 37303-22 (18/171705, 18/212469, 18/361196, 18/574635, 18/791095, 679/022509, 679/042417, 679/112224, 679/112340), 37303-24 (18/575639, 18/741572, 26/F66915, 679/022510, 679/112226, 679/122313, 679/122333), 37303-26 (18/152346, 18/214471, 26/F66916, 679/012403, 679/112230, 679/122323), 37303-28 (26/F66917, 679/112231, 679/122328), 37303-30 (18/192716, 18/591743, 679/122334). 2. Article Number (Lot Numbers): 32271-10 (18/061949, 18/231167, 679/062351), 32271-12 (18/048927, 26/F66235, 679/042215, 679/062342, 679/062405, 679/062489, 679/072271), 32271-14 (18/049928, 18/231165, 26/F66140, 679/042210, 679/052340, 679/052518, 679/062471, 679/062490, 679/072267), 32271-16 (18/231164, 18/311125, 26/F66141, 679/042216, 679/052519, 679/062315, 679/062483, 679/072272, 679/072401), 32271-18 (18/062950, 26/F65556, 26/F66236, 679/032261, 679/062332, 679/072402), 32271-20 (18/063951, 26/F66237, 679/052326), 32271-22 (18/064952, 26/F65557, 679/032263, 679/062354), 32271-24 (18/065953, 26/F65558, 679/062333), 32271-26 (679/032265), 32271-28 (26/F65559, 679/062357, 679/052520), 32271-32 (26/F65561), 32271-34 (26/F65562). 3. Article Number (Lot Numbers): 37304-10 (18/635486, 26/F24799, 679/012419, 86/011176), 37304-12 (18/061352, 18/521172, 18/5211721, 18/576643, 18/630487, 679/012433, 679/092207, 679/122409), 37304-14 (18/081360, 18/531173, 18/5311731, 18/577644, 18/631488, 679/012434, 679/092208, 679/122410), 37304-16 (18/541174, 18/578645, 18/632489, 679/012444, 679/092201, 679/122411), 37304-18 (18/092020, 18/551175, 18/633490, 679/022403, 679/092202), 37304-20 (18/011320, 18/561176, 18/634491, 679/012287, 679/012420, 679/082217, 86/483331), 37304-22 (18/241273, 679/012288, 679/012421), 37304-24 (18/021321, 18/571177, 679/012289, 679/012445). 4. Article Number (Lot Numbers): 21413-6 (26/F52124). 5. Article Number (Lot Numbers): 21431-ME (679/102045). 6. Article Number (Lot Numbers): 21482 (679/092020), 21483 (26/F55902). 7. Article Number (Lot Numbers): 21421-5 (26/F65377, 679/092039). 8. Article Number (Lot Numbers): 37241-10 (18/012169, 18/610104, 18/701013), 37241-12 (18/231295, 18/767108), 37241-14 (18/511016, 18/537392, 26/F24807, 679/012280), 37241-16 (18/005340, 18/121106, 18/231387, 18/521017, 18/560517), 37241-18 (18/232328, 18/531018, 679/072408, 86/311848), 37241-20 (18/233388, 18/541020, 18/562519, 679/072419, 86/287847), 37241-22 (18/066109, 18/234389, 18/551021, 26/F24594, 26/F24811), 37241-24 (18/561022, 18/564521, 18/608110, 26/F24595), 37241-26 (18/611111, 18/711023, 26/F24596, 26/F24813), 37241-28 (18/612112, 18/721024, 18/811597, 26/F24597, 86/485508), 37241-30 (18/015172, 18/613113, 18/731019, 26/F24598). 9. Article Number (Lot Numbers): 21015-08 (00/0123-12, 00/0321-43, 00/0723-07), 21015-09 (00/0123-13).
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
15,517 units