FDA Enforcement
Class II
Terminated
Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000
Recall: Z-1602-2020
·
Reported April 8, 2020
Enforcement
- Recall Number
- Z-1602-2020
- Event ID
- 85083
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Stryker Communications
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- April 8, 2020
- Initiation Date
- October 11, 2019
- Classification Date
- March 27, 2020
- Termination Date
- October 18, 2022
- Address
- 571 Silveron Blvd, N/A, Flower Mound, TX, 75028, United States
Description
Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000
Reason
The firm has become aware that there is a likelihood that a missing snap ring or spanner nut may result in an insufficient mount force that causes the surgical light to fall. If the suspension of the surgical light falls, it may potentially cause serious injuries to Health Care Providers and/or patients.
Code Info
Serial Number Range: 091316 70016 through 083117-73252
Distribution
US: Nationwide OUS: Canada
Quantity
1415 Surgical Lights