FDA Enforcement Class II Terminated

Vitek 2 AST-YS07 REF 414 967, Fungal Susceptibility Card, 20 cards per carton, IVD. Intended for use with the Vitek 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant yeasts to antifungal agents when used as instructed.

Recall: Z-1602-2015 · Reported May 20, 2015

Enforcement

Recall Number
Z-1602-2015
Event ID
70868
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Biomerieux Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
May 20, 2015
Initiation Date
March 27, 2015
Classification Date
May 12, 2015
Termination Date
March 24, 2017
Address
595 Anglum Rd, N/A, Hazelwood, MO, 63042-2320, United States

Description

Vitek 2 AST-YS07 REF 414 967, Fungal Susceptibility Card, 20 cards per carton, IVD. Intended for use with the Vitek 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant yeasts to antifungal agents when used as instructed.

Reason

The test kits have Out-of-Range-Low (OORL) Flucytosine QC results for Candida krusei ATCC 6258

Code Info

lot 287303120, 02-APR-15; lot 287303122, 02-APR-15; lot 287306710, 08-MAY-15; lot 287308210, 23-MAY-15; lot 287311710, 27-JUN-15; lot 287314910, 29-JUL-15; lot 287314912, 29-JUL-15; lot 287315710, 06-AUG-15; lot 287317310, 22-AUG-15; lot 287317710, 26-AUG-15; lot 287318920, 07-SEP-15; lot 287319110, 09-SEP-15; lot 287319820, 16-SEP-15; lot 287321940, 07-OCT-15; lot 287322540, 13-OCT-15; lot 287323520, 23-OCT-15; lot 287324220, 30-OCT-15; lot 287324240, 30-OCT-15; lot 287326220, 19-NOV-15; lot 287327620, 03-DEC-15; lot 287329620, 23-DEC-15; lot 287330520, 01-JAN-16; lot 287331920, 15-JAN-16; lot 287332520, 21-JAN-16; lot 287332522, 21-JAN-16; lot 287334810, 13-FEB-16; lot 287335420, 19-FEB-16; lot 287336220, 27-FEB-16; lot 287337620, 12-MAR-16; lot 287338020, 16-MAR-16; lot 287340140, 06-APR-16; lot 287341820, 23-APR-16; lot 287341822, 23-APR-16; lot 287344620, 21-MAY-16; lot 287344622, 21-MAY-16; lot 287346710, 11-JUN-16; lot 287346720, 11-JUN-16; lot 287350120, 15-JUL-16; and lot 287350122, 15-JUL-16.

Distribution

Distributed only to foreign subsidiaries and distributors, including Canada and Mexico.

Quantity