FDA Enforcement Class II Terminated

Syngo Dynamics; Kinetdx Picture Archiving and Communications System

Recall: Z-1601-2016 · Reported May 11, 2016

Enforcement

Recall Number
Z-1601-2016
Event ID
73907
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 11, 2016
Initiation Date
April 14, 2016
Classification Date
May 4, 2016
Termination Date
March 20, 2017
Address
40 Liberty Blvd, Malvern, PA, 19355-1418, United States

Description

Syngo Dynamics; Kinetdx Picture Archiving and Communications System

Reason

Siemens has identified changes in default carry forward workflow using VA10B_HF03 version of Syngo Dynamics. If the template design has measurements inside a carry forward-configured report section, this will result in carrying forward the entire section of the report including data from previous studies. This may result in the physician making analysis based on the old measurements.

Code Info

Model Numbers: 10091637, 10091673, 10091805, 10091807 Serial Numbers: 85021, 85040, 85042, 85045, 85047, 85057, 85079, 85080, 85081, 85086, 85090, 85115, 85143, 85145, 85146, 85147, 85149, 85157, 85167, 85168, 85170, 85172, 85177, 85178, 85185, 85186, 85189, 85191, 85234, 85235, 85239, 85281, 85307, 85309, 85322, 85325, 85326, 85329, 85335, 85338, 85350, 85353, 85354, 85368, 85370, 85374, 85377, 85380, 85385, 85387, 85401, 85418, 85421, 85422, 85427, 85435, 85442, 85447, 85462, 85467, 85468, 85469, 85470, 85473, 85481, 85482, 85484, 85859, 85860, 85866, 85867, 85871, 85872, 85873, 85874, 85875, 85876, 85880, 85883, 85884, 85893, 85896, 85910, 85918

Distribution

Nationwide: AZ, CA, IL, IN, MA, MD, MI, MN, NJ, NY, OH, OR, PA, TN, TX, UT, WA, WI

Quantity

84 units