FDA Enforcement
Class II
Ongoing
Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
Recall: Z-1600-2026
·
Reported April 1, 2026
Enforcement
- Recall Number
- Z-1600-2026
- Event ID
- 98341
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- I.T.S. GmbH
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 1, 2026
- Initiation Date
- January 23, 2026
- Classification Date
- March 20, 2026
- Address
- Autal 28, Lasnitzhohe, N/A, Austria
Description
Pelvic Reconstruction System (PRS) with the below descriptions and corresponding article numbers. 1. Fixation Screw; Article Number: 70312. 2. Spike short; Article Number: 70314.
Reason
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Code Info
1. Article Number (Lot Number): 70312 (26/F65133). 2. Article Number (Lot Number): 70314 (86/619736).
Distribution
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
Quantity
11 units