I.T.S. Screw System with the below descriptions and corresponding article numbers. 1. Cancellous Screw, Cannulated; Article Numbers: 31404-16, 31404-18, 31404-20, 31404-22, 31404-24, 31404-26, 31404-28, 31404-30, 31404-32, 31404-34, 31404-36, 31404-38, 31404-40, 31404-42, 31404-44, 31404-46, 31404-48, 31404-50, 31404-55, 31404-60, 31404-65, 31404-70, 31404-75, 31404-80, 31457-30, 31457-36, 31457-40, 31457-46, 31457-50, 31457-55, 31457-60, 31652-100, 31652-105, 31652-110, 31652-115, 31652-120, 31652-25, 31652-30, 31652-35, 31652-40, 31652-45, 31652-55, 31652-60, 31652-65, 31652-70, 31652-75, 31652-80, 31652-85, 31652-90, 31652-95, 31731-100, 31731-105, 31731-110, 31731-115, 31731-120, 31731-50, 31731-60, 31731-65, 31731-70, 31731-75, 31731-80, 31731-85, 31731-90, 31731-95, 31732-60, 31734-100, 31734-105, 31734-110, 31734-115, 31734-120, 31734-45, 31734-50, 31734-60, 31734-65, 31734-70, 31734-75, 31734-80, 31734-85, 31734-90, 31734-95. 2. Cortical Screw, Cannulated; Article Numbers: 31357-44, 31357-22, 31357-48, 31277-18, 31357-20, 31357-22, 31357-24, 31357-26, 31357-28, 31357-30, 31357-32, 31357-34, 31357-36, 31357-38, 31357-40, 31357-42, 31357-44, 31357-46, 31357-48, 31357-60, 31357-65, 31357-70. 3. Washer; Article Number: 36851. 4. Washer, Concave; Article Numbers: 36432, 36653, 36732. 5. Washer, Flat; Article Numbers: 36431, 36651, 36653, 36731.
Enforcement
- Recall Number
- Z-1599-2026
- Event ID
- 98341
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- I.T.S. GmbH
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 1, 2026
- Initiation Date
- January 23, 2026
- Classification Date
- March 20, 2026
- Address
- Autal 28, Lasnitzhohe, N/A, Austria
Description
I.T.S. Screw System with the below descriptions and corresponding article numbers. 1. Cancellous Screw, Cannulated; Article Numbers: 31404-16, 31404-18, 31404-20, 31404-22, 31404-24, 31404-26, 31404-28, 31404-30, 31404-32, 31404-34, 31404-36, 31404-38, 31404-40, 31404-42, 31404-44, 31404-46, 31404-48, 31404-50, 31404-55, 31404-60, 31404-65, 31404-70, 31404-75, 31404-80, 31457-30, 31457-36, 31457-40, 31457-46, 31457-50, 31457-55, 31457-60, 31652-100, 31652-105, 31652-110, 31652-115, 31652-120, 31652-25, 31652-30, 31652-35, 31652-40, 31652-45, 31652-55, 31652-60, 31652-65, 31652-70, 31652-75, 31652-80, 31652-85, 31652-90, 31652-95, 31731-100, 31731-105, 31731-110, 31731-115, 31731-120, 31731-50, 31731-60, 31731-65, 31731-70, 31731-75, 31731-80, 31731-85, 31731-90, 31731-95, 31732-60, 31734-100, 31734-105, 31734-110, 31734-115, 31734-120, 31734-45, 31734-50, 31734-60, 31734-65, 31734-70, 31734-75, 31734-80, 31734-85, 31734-90, 31734-95. 2. Cortical Screw, Cannulated; Article Numbers: 31357-44, 31357-22, 31357-48, 31277-18, 31357-20, 31357-22, 31357-24, 31357-26, 31357-28, 31357-30, 31357-32, 31357-34, 31357-36, 31357-38, 31357-40, 31357-42, 31357-44, 31357-46, 31357-48, 31357-60, 31357-65, 31357-70. 3. Washer; Article Number: 36851. 4. Washer, Concave; Article Numbers: 36432, 36653, 36732. 5. Washer, Flat; Article Numbers: 36431, 36651, 36653, 36731.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
1. Article Number (Lot Numbers): 31404-16 (679/062339, 679/122213, 679/122332), 31404-18 (679/122212, 679/062340), 31404-20 (18/089221, 679/062343, 679/122241), 31404-22 (18/061535, 679/012416, 679/062350), 31404-24 (18/071536, 679/062362, 679/122246), 31404-26 (679/122245, 679/062363), 31404-28 (18/186278, 679/062364), 31404-30 (18/187299, 679/012401, 679/062365, 679/122242), 31404-32 (18/111540, 679/062366, 679/122211), 31404-34 (679/122206, 18/121544, 679/012402, 679/042414, 679/072302, 679/122206), 31404-36 (18/188300, 679/052403, 679/072305, 679/112223), 31404-38 (18/189301, 679/112213, 679/012455, 679/072309), 31404-40 (18/181043, 679/062321, 679/122201, 679/122310, 679/052404, 679/032532), 31404-42 (679/122203, 18/191303, 679/072310, 679/012418, 679/052424, 679/052505), 31404-44 (18/192304, 679/112232, 679/052332, 679/052425, 679/052508), 31404-46 (679/122240, 679/062322, 679/112210, 679/052426, 679/042505), 31404-48 (679/052427, 679/072317, 679/122248), 31404-50 (18/194306, 679/112228, 679/062301, 679/122311, 679/052406, 679/052413, 679/042506, 679/062501), 31404-55 (18/195307, 679/122258, 18/211647, 679/072326, 679/012452, 679/062467, 679/062511), 31404-60 (18/196308, 679/112240, 679/052325, 679/122312, 679/062485, 679/062512), 31404-65 (679/122259, 679/072327, 18/205310, 679/022410), 31404-70 (679/012301, 18/197309, 679/072328, 679/022413, 679/062479), 31404-75 (18/205411, 18/381989, 679/012302), 31404-80 (18/172638, 18/205232, 18/321164, 679/022416), 31457-30 (26/F50332), 31457-36 (18/198343), 31457-40 (18/199344), 31457-46 (18/947783, 18/736681), 31457-50 (18/201346, 26/F67796), 31457-55 (18/202347, 26/F67797), 31457-60 (18/203355), 31652-100 (679/122117, 679/022350, 679/102335, 679/022335, 18/924999), 31652-105 (679/022336, 679/032377), 31652-110 (679/032380, 679/102339), 31652-115 (679/102336), 31652-120 (679/092002, 679/102347), 31652-25 (679/112134, 679/032340), 31652-30 (679/032341), 31652-35 (679/082315), 31652-40 (679/082316), 31652-45 (679/032330, 679/092332), 31652-55 (679/032391, 679/092333), 31652-60 (679/022337, 679/032378, 679/092336), 31652-65 (679/022338, 679/032322, 679/092347), 31652-70 (679/032320, 679/032365, 679/042405, 679/112154), 31652-75 (26/F65330, 679/032348, 679/032386), 31652-80 (679/112181, 679/032321, 679/092350, 679/032362, 679/092348, 679/042410, 18/920995), 31652-85 (679/122145, 679/032382, 679/022349, 679/102301, 679/102305, 679/042411, 18/541045), 31652-90 (679/012201, 679/032383, 679/032383, 679/022347, 679/042406), 31652-95 (18/923998, 679/012202, 679/032316), 31731-100 (18/775943, 26/F58803, 679/102420), 31731-105 (18/776944, 26/F58804, 26/F65649, 679/102421), 31731-110 (26/F58805, 26/F65650, 679/102422), 31731-115 (26/F58806, 26/F65651, 679/102423), 31731-120 (26/F58807, 26/F65652, 679/102436), 31731-50 (09/1244, 26/F14488, 26/F21185, 26/F58793), 31731-60 (26/F58795, 26/F65655), 31731-65 (26/F65656, 26/F58796), 31731-70 (26/F58797, 26/F65657), 31731-75 (26/F58798, 26/F65658), 31731-80 (679/072040, 26/F58799, 26/F65659), 31731-85 (18/331317, 26/F65660, 679/072044), 31731-90 (09/1247, 09/1248, 26/F58801, 26/F65661, 46/612), 31731-95 (18/581247, 26/F58802, 26/F65662, 679/102437), 31732-60 (18/961265), 31734-100 (26/F65672), 31734-105 (26/F58839, 26/F65673), 31734-110 (26/F58840, 26/F65674, 679/032425), 31734-115 (26/F65675), 31734-120 (26/F58842, 26/F65676, 26/F72277), 31734-45 (18/761192), 31734-50 (26/F58828, 26/F65680), 31734-60 (26/F58830), 31734-65 (26/F58831, 26/F65683, 679/032428), 31734-70 (26/F58832, 26/F65684, 26/F72275, 679/032429), 31734-75 (26/F58833, 26/F65685), 31734-80 (26/F58834, 26/F65686, 679/032435), 31734-85 (26/F58835, 26/F65687, 26/F72276), 31734-90 (26/F65688, 679/032436), 31734-95 (26/F58837, 26/F65689, 679/042401). 2. Article Number (Lot Numbers): 31277-18 (26/F36049), 31357-20 (18/091247), 31357-22 (679/031707), 31357-24 (86/485378), 31357-26 (18/121250), 31357-28 (26/F65909), 31357-30 (18/165201, 18/341054), 31357-32 (18/166217), 31357-34 (18/167218), 31357-36 (18/168219, 26/F65910), 31357-38 (18/167220), 31357-40 (26/F65911, 18/170221), 31357-42 (26/F65912), 31357-44 (18/171222, 18/361059, 26/F65913), 31357-46 (18/861060), 31357-48 (18/172223, 26/F65914), 31357-60 (26/F65648), 31357-65 (679/012125), 31357-70 (18/902013). 3. Article Number (Lot Numbers): 36851 (26/F59918, 26/F61351, 26/F64885, 26/F64886, 26/F65908, 26/F66506, 26/F71413, 26/F71764, 26/F72364, 26/F73020, 26/F74288). 4. Article Number (Lot Numbers): 36432 (26/F65379, 26/F67799, 26/F72744, 774/300002), 36653 (26/F62590, 26/F68455), 36732 (26/F67800). 5. Article Number (Lot Numbers): 36431 (26/F62589, 26/F65378, 26/F66505, 26/F67798, 26/F70092, 26/F72743, 26/F73017, 26/F73494), 36651 (26/F61694, 26/F65380, 26/F68454, 26/F73018), 36653 (26/F73019), 36731 (26/F60987, 26/F62591, 26/F67523, 26/F70093, 26/F72363, 26/F73495).
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
5,568 units