FDA Enforcement Class II Ongoing

Medtronic, REF: MMT-7008A, Enlite Glucose Sensor for the following packaging: ENLITE PLUS 5PK 23L 1CL /GTIN: 20643169541706, ENLT PLUS 5PK 2L 1CL PR/ GTIN: 20643169786725, ENLITE PLUS 5PK 23L 1CL/ GTIN: 20763000421292 - Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.

Recall: Z-1599-2021 · Reported May 26, 2021

Enforcement

Recall Number
Z-1599-2021
Event ID
87476
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Medtronic Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
May 26, 2021
Initiation Date
February 26, 2021
Classification Date
May 17, 2021
Address
18000 Devonshire St, N/A, Northridge, CA, 91325-1219, United States

Description

Medtronic, REF: MMT-7008A, Enlite Glucose Sensor for the following packaging: ENLITE PLUS 5PK 23L 1CL /GTIN: 20643169541706, ENLT PLUS 5PK 2L 1CL PR/ GTIN: 20643169786725, ENLITE PLUS 5PK 23L 1CL/ GTIN: 20763000421292 - Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.

Reason

Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.

Code Info

All Lot numbers within active shelf life

Distribution

Worldwide distribution.

Quantity

213,110 units