FDA Enforcement Class II Ongoing

Strep test: a. McKesson Medical-Surgical McKesson Rapid Test Kit McKesson Consult Infectious Disease Immunoassay Strep A Test Throat / Tonsil Saliva Sample 25 Tests, Model Number: 5003. b. QUIDEL Sofia TEST KIT, STREP A+ SOFIA (25/KT 12KT/CS) Model Number: 20274. c. Alere SAN DIEGO INC McKesson TEST, KIT STREP A DIPSTICK WAIVED (50/KT) Exempt Catalog #4999. d. In-Line Control Swab Quickvue In-Line Strep A Test Positive Level / Negative Level 12 Swabs (IVD) Catalog # 00345. e. (added 11/30/22) Abbott Rapid Test Kit Acceava Infectious Disease Immunoassay Strep A Test Throat Swab Sample 50 Tests, Model Number: 92001

Recall: Z-1598-2022 · Reported August 31, 2022

Enforcement

Recall Number
Z-1598-2022
Event ID
90336
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Mckesson Medical-Surgical Inc. Corporate Office
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Report Date
August 31, 2022
Initiation Date
May 25, 2022
Classification Date
August 25, 2022
Address
9954 Maryland Drive, Deep Run Iii Ste. 4000, Richmond, VA, 23233, United States

Description

Strep test: a. McKesson Medical-Surgical McKesson Rapid Test Kit McKesson Consult Infectious Disease Immunoassay Strep A Test Throat / Tonsil Saliva Sample 25 Tests, Model Number: 5003. b. QUIDEL Sofia TEST KIT, STREP A+ SOFIA (25/KT 12KT/CS) Model Number: 20274. c. Alere SAN DIEGO INC McKesson TEST, KIT STREP A DIPSTICK WAIVED (50/KT) Exempt Catalog #4999. d. In-Line Control Swab Quickvue In-Line Strep A Test Positive Level / Negative Level 12 Swabs (IVD) Catalog # 00345. e. (added 11/30/22) Abbott Rapid Test Kit Acceava Infectious Disease Immunoassay Strep A Test Throat Swab Sample 50 Tests, Model Number: 92001

Reason

Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

Code Info

GTIN: a. 00612479240100; 00612479202733. b. 30014613202745. c. 00612479202726; 612479240094. d. 30014613003458. e. 10815845020031, 00815845020034.

Distribution

US Nationwide

Quantity

a. 214 kits b. 1cs c. 53 kits d. 4 bx e. 1 KT