FDA Enforcement
Class II
Terminated
MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.
Recall: Z-1598-2016
·
Reported May 11, 2016
Enforcement
- Recall Number
- Z-1598-2016
- Event ID
- 73947
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Siemens Medical Solutions USA, Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 11, 2016
- Initiation Date
- April 13, 2016
- Classification Date
- May 3, 2016
- Termination Date
- April 17, 2017
- Address
- 40 Liberty Blvd, Malvern, PA, 19355-1418, United States
Description
MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.
Reason
The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.
Code Info
Model # 10432914 and 10432915 Serial Numbers: 42472, 42489, 42477, 46169, 46178, 46175, 42478, 42485
Distribution
US nationwide distribution to OH, FL, TN, MN, OH, NY, VA, and MN.
Quantity
8 systems