FDA Enforcement Class II Terminated

MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.

Recall: Z-1598-2016 · Reported May 11, 2016

Enforcement

Recall Number
Z-1598-2016
Event ID
73947
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Siemens Medical Solutions USA, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 11, 2016
Initiation Date
April 13, 2016
Classification Date
May 3, 2016
Termination Date
April 17, 2017
Address
40 Liberty Blvd, Malvern, PA, 19355-1418, United States

Description

MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.

Reason

The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.

Code Info

Model # 10432914 and 10432915 Serial Numbers: 42472, 42489, 42477, 46169, 46178, 46175, 42478, 42485

Distribution

US nationwide distribution to OH, FL, TN, MN, OH, NY, VA, and MN.

Quantity

8 systems