FDA Enforcement Class II Terminated

Nuclear Gamma Cameras (ADAC Vertex Classic, ADAC VERTEX PLUS/SOLUS/CARDIO, ADAC VERTEX V60) Philips Medical Systems, Cleveland, OH

Recall: Z-1598-2015 · Reported May 20, 2015

Enforcement

Recall Number
Z-1598-2015
Event ID
71190
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 20, 2015
Initiation Date
March 13, 2015
Classification Date
May 12, 2015
Termination Date
June 26, 2018
Address
595 Miner Rd, N/A, Cleveland, OH, 44143-2131, United States

Description

Nuclear Gamma Cameras (ADAC Vertex Classic, ADAC VERTEX PLUS/SOLUS/CARDIO, ADAC VERTEX V60) Philips Medical Systems, Cleveland, OH

Reason

During clinical use of a Vertex Plus Gamma Camera, as the detector heads were in the relative 180 degree position, Detector #1 drifted down and made contact with the patients chin during a whole body study acquisition.

Code Info

Vertex Classic; Model #: 890122; Serial #s: 9404100, 9601238, 9609084, 0407V, 0682V, 0795V, 0822VP2, 103, 20, 500, 5405V, 64, 9308007, 9309020, 9310032, 93-10660-1292, 9401059, 9401061, 9403076, 9403083, 9404106, 9407122, 9409148, 9410002, 9411006, 9411007, 9411016, 9501049, 9501210, 9502067, 9502086, 9503003, 9503079, 9503109, 9504019, 9505021, 9505124, 9506140, 9508169, 9508171, 9508184, 9508193, 9509198, 9509206, 9510222, 9510226, 9601146, 9602249, 9608042, 9901004, B0106154, B02120033, M@10229AMT, M@207JD, M@4404V, M@4431V, M@4432V, M@4480V, M@4623V, M@4630V, M@4639V, M@4686V, M@4692V, M@4858V, M@4893V, M@4959V, M@5032V, M@5033V, M@5038V, M@5105V, M@5145V, M@5158V, M@5189V, M@5280V, M@5289V, M@5338V, M@5360V, M@5373V, M@5398V, M@5407V, M@5412V, M@5431V, M@5439V, M@5467V, M@5476V, M@5497V, M@5498V, M@5515V, M@5526V, M@5590V, M@5617V, M@5627V, M@5661V, M@5662V, M@5663V, M@5665V, M@5681V, M@5688V, M@5689V, M@5751V, M@5752V, M@5782V, M@5811V, M@5835V, M@5869V, M@5900V, M@5913V, M@5915V, M@5960V, M@5970V, M@6004V, M@6007V, M@6010V, M@6056V, M@6900V, M@9304RV, M@9645RV, M@MVS5236V, M@MVS5852V, M@O1325, UNK-AMT1238A. VERTEX PLUS/SOLUS/CARDIO: Model #: 88297; Serial #s: 9603015, 9605036, 9610076, 9610077, 9703171, 9706192, 9710103, 9711277, 9802304, 9803109, 9807121, 9808122, 9809056, 9809124, 9809415, 9809424, 9912461, _0002032, ~9608052, 0007001M, 0008009B, 0011002B, 0102019V, 0103020V, 01069VM, 01070VM, 0111001V, 01295VM, 01296K, 01864VM, 03026VM, 0939VM, 100098-01N, 1006, 1008, 10252, 1027, 1037, 10630, 1092, 1097, 11, 130234-01NVERTEX, 2035, 2038, 2047, 3059, 3060, 3064, 4007, 4078, 5934V, 6052, 6261, 7373VM, 7376VM, 7386VM, 7560VM, 7561VM, 8002, 802080322, 85247, 9054VM/V, 9056VM/V, 9073152, 9130VM, 9205345, 940384, 9408136, 9503084, 9504096, 9506016, 9506145, 9507163-CN, 9511123, 9512008, 9512009, 9512018, 9601015, 9601036, 9601240-CN, 9602161, 9602245, 9602246, 9602254, 9603336, 9604016, 9604021, 9605026, 9605037, 9606211, 9607015, 9607024, 9608046, 9608049, 9608050, 9608052, 9608064, 9609063, 9609081, 9610058, 9611003, 9611077, 9611082, 9612072, 9701127, 9702061-CN, 9702095, 9702150, 9703154, 9704044, 9704178, 9704183, 9705004, 9705186, 9706005, 9706182, 9706189, 9707002, 9707047, 9707050, 9707054, 9708062, 9708065, 9708068-CN, 9708069, 9708076, 9708078, 9708084, 9708227, 9709022, 9710121, 9710127, 9710244, 9711025, 9711034, 9711036, 9711074, 9711078, 9711261, 9711271, 9801047, 9801227, 9801232, 9801297, 9801299-CN, 9802050, 9802255, 9802260, 9802300, 9802306, 9802397, 9803003, 9803014, 9803067, 9803083, 9803269, 9803272, 9803288, 9803290, 9803293, 9803295, 9803312, 9803316, 9803322, 9804069, 9804330, 9804333, 9804342, 9805344, 9805350, 9806006, 9806014, 9806363, 9806377, 9806388, 98066002, 9807002, 9807391, 9807421, 9808036, 9808090, 9808108, 9808451, 9808452, 9808457, 9808480, 9808829, 9809062 (-CN), 9809065-CN, 9809067, 9809113, 9809122, 9809125, 9809498, 9809687, 9810075, 9810136, 9810428ADAC#7973VM, 9810442, 9810551, 9810557, 9811020, 9811093, 9811569, 9811737, 9812582, 9901597, 9901605, 9902150, 9902622, 9902630, 9903632, 9903637, 9903641, 9904089, 9905012, 9905659, 9905667, 9906670, 9907016, 9907017, 9907019, 9907029, 9909161, 9910031, 9911458, 9911698, 9911702, 9912181, 9912463, AD1261, AMT00848VM, B0006129, B0007273, B0008003, B0008012, B0008021, B0008030, B001200212, B0105138, B0106169, B01070191, B01070192, B0107179, B01080204, B01080207, B01090227, B01100231, B01100233, B01100238, B01110251, B02050071, B02050262, B02060167, B02060270, B02060279, B0206027S, B02070290, B02070300, B0207296, B02080318, B0208312, B02090327, B02090333, B02090334, B02100345, B03010355, B03010359, B03020048, B03080380, C0008070, C01080152, C01080154, C03050230, C03060193, M@10458, M@10573, M@10582, M@10589, M@10772, M@10837, M@10850, M@10927, M@10967, M@11020, M@11036, M@11109, M@11140, M@11178, M@11498, M@11525, M@11571, M@11590, M@11881, M@11949, M@11957, M@11972, M@12219, M@12412, M@12591, M@12711, M@12787, M@13010, M@13296, M@13466, M@13498, M@13543, M@13549, M@13665, M@13688, M@13691, M@13748, M@13813, M@22752, M@28051, M@4750019856, M@5828J, M@5906N, M@5919J, M@5936N, M@5956J, M@5961K, M@6002J, M@6012J, M@6075VM, M@6108VA, M@6123VA, M@6160VM, M@6183VA, M@6258J, M@6323VM, M@6324J, M@6330VA, M@6341VA, M@6345VA, M@6346VM, M@6348VM, M@6349VM, M@6365VM, M@6368J, M@6370VA, M@6375J, M@6376J, M@6460VM, M@6464VM, M@6470VA, M@6481VA, M@6496VM, M@6528VA, M@6541VA, M@6542VM, M@6566VM, M@6573VM, M@6596VM, M@6600VA, M@6638VM, M@6639J, M@6675VA, M@6716VM, M@6717VM, M@6762VM, M@6773VM, M@6776VM, M@6797K, M@6805VM, M@6823VM, M@6840K, M@6841VM, M@6867VM, M@6877VM, M@6886VM, M@6891VA, M@6924VM, M@6931VM, M@6932VM, M@6943VM, M@6946VM, M@6947J, M@6985J, M@7003VA, M@7004VA, M@7005VA, M@7017VM, M@7024VM, M@7034K, M@7042VM, M@7053VM, M@7073VM, M@7085K, M@7086J, M@7095VM, M@7100VM, M@7101VM, M@7102VM, M@7151VM, M@7153VM, M@7159VM, M@7166K, M@7171VM, M@7172K, M@7181J, M@7190VM, M@7198VA, M@7215K, M@7216VM, M@7217VM, M@7221VA, M@7222VA, M@7229VM, M@7233VM, M@7237VM, M@7245J, M@7246VM, M@7247VM, M@7250VM, M@7300J, M@7308VM, M@7376VM, M@7380VM, M@7384VM, M@7390VM, M@7393VM, M@7394VM, M@7410VM, M@7420VM, M@7464VA, M@7470VM, M@7490VM, M@7494K, M@7501VM, M@7504VM, M@7505VM1, M@7514VM, M@7515VM, M@7536VM, M@7537VM, M@7538VM, M@7556VM, M@7557VM, M@7569VM, M@7572VM, M@7591VM, M@7592VM, M@7603K, M@7605VM, M@7606VM, M@7623K, M@7626K, M@7627K, M@7628VA, M@7638VM, M@7643VM, M@7654VM, M@7664VM, M@7665K, M@7673VM, M@7674K, M@7679VM, M@7682K, M@7685VM, M@7691VM, M@7695VM, M@7697VM, M@7723VM, M@7729VM, M@7733VM, M@7739K, M@7760VM, M@7761VM, M@7770VM, M@7776VM, M@7789K, M@7801VM, M@7810VM, M@7811VM, M@7812VM, M@7819VA, M@7823K, M@7824VM, M@7830K, M@7855J, M@7857VM, M@7860VM, M@7861VM, M@7874VM, M@7877VM, M@7918K, M@7924K, M@7925K, M@7929VM, M@7930VM, M@7932VM, M@7946J, M@7957VM, M@7961VM, M@7962VM, M@7965VM, M@7968VM, M@7975K, M@7987VM, M@7995VM, M@7997VM, M@8004VM, M@8013VM, M@8018J, M@8021VM, M@8044VM, M@8060VM, M@8064VM, M@8076K, M@8089VM, M@8096VM, M@8105VM, M@8124VA, M@8124VA, M@8146VM, M@8155VM, M@8178VM, M@8203VM, M@8219VA, M@8220K, M@8240K, M@8241VM, M@8248VM, M@8252VM, M@8284K, M@8301K, M@8302VM, M@8306K, M@8337VM, M@8356K, M@8380VM, M@8399VM, M@8484VM, M@8528VM, M@8537VM, M@8551VM, M@8561K, M@8606K, M@8609VM, M@8651VM, M@8683VM, M@8686K, M@8693VM, M@8726VM, M@8738K, M@8759K, M@8771VM, M@8781VM, M@8797K, M@8799RV, M@8800VM, M@8803K, M@8812VM, M@8814VM, M@8882K, M@8888K, M@8897K, M@8903K, M@8906RK, M@8920VM, M@8922VM, M@8935K, M@8948RV, M@8955K, M@8990K, M@8996VM, M@9032K, M@9046K, M@9049VM, M@9081K, M@9091VM, M@9107RV, M@9166VM, M@9185K, M@9194VM, M@9221VM, M@9222VM, M@9247RV, M@9252VM, M@9267VM, M@9309RV, M@9316VM, M@9327K, M@9338VM, M@9340VM, M@9360VM, M@9369VM, M@9418VM, M@9426K, M@9438K, M@9471RV, M@9515VM, M@9532VM, M@9542VM, M@9543K, M@9583VM, M@9607VM, M@9653VM, M@9672VM, M@9724VM, M@9730K, M@9816K, M@9831VM, M@AMT1258, M@MCD6416VM, M@MCD7036VM, M@R372820, M@S5195VA, M@S5365-V, M@S5820VM, M@S5907V11, M@S6155VA, M@S7156VM, M@S7223K, M@S7559VA, M@S7571VM, M@S7856K, M@S7864K, M@S7904VM, M@S7941VA, M@S8687K, M0101023, R03010363, S0011003, S6235VM, STOR_QUO_47521360, UNK-6597VM, UNK-7590VM, UNK-8533VM1, UNK-8860VM, UNK-9868K, V0007070, V0010090, V0012099, V0104109, V0104110, V0105112, V0106121, V0106123, V0106124, V01070133, V01090146, V01100149, V01100151, V01100156, V02050166, V02070174, V02070179, V02080190, V02080194, V02080199, V02080200, V02090206, V03060224, V03070229, V03091241. VERTEX V60 Model #: 882030; Serial #s: 9601243, 9602253, 9603008, 9603013, 9604028, 9604041, 9605033, 9606038, 9610256, 9704175, 9704177, 9801284, 9803324, 9807089, 9810434, 9904005, 970001-9610187, M0007018, V0009076, V0011096, V0106122, V0106125, V02050163, V02050165, V02070181, 9705007, 9711262, 00559VM, 00634V, 00657VA, 00970VA, 01190VM, 01697VM, 01894VM, 0371V, 10588, 1189-17, 12001, 1273-06, 1274-06, 9601004, 9607063, 9707206, 9710247, 9712001, 9712008, 9712009, 9803003, 9806019, 9808021, 9902027, 9909039, M0007001-CN, V0005062, V01090141, V01090142, V03080235, V03080236, V2070177, VETEJ, X0005002, X0010004, X0101006, X0101007, X03030012, Z00657VA, Z01798VM.

Distribution

Worldwide Distribution - US (nationwide) including Puerto Rico and to the states of : AL, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI and WV., and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Costa Rica, Denmark, Egypt, France, Germany, Ireland, Italy, Japan, Lebanon, Malaysia, Netherlands, Norway, Pakistan, Portugal, Republic of Korea, Russia, South Korea, Spain, Sweden, Syrian Arab Republic, Thailand and United Kingdom.

Quantity

918 units