FDA Enforcement Class II Ongoing

DxU 850m Iris Urine Microscopy Analyzer Catalog Number C49513, in-vitro diagnostic device, urine analysis

Recall: Z-1597-2023 · Reported May 24, 2023

Enforcement

Recall Number
Z-1597-2023
Event ID
92056
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 24, 2023
Initiation Date
April 5, 2023
Classification Date
May 17, 2023
Address
11800 Sw 147th Ave, N/A, Miami, FL, 33196-2500, United States

Description

DxU 850m Iris Urine Microscopy Analyzer Catalog Number C49513, in-vitro diagnostic device, urine analysis

Reason

Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.

Code Info

UDI/DI 15099590748647, All analyzers with Windows 10, Software Versions 8.5, 8.5.1 and 8.6

Distribution

Worldwide Distribution

Quantity

107 units