Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6, 21094-8. 3. Pilon Plate, Small; Article Numbers: 21093-4, 21093-6.
Enforcement
- Recall Number
- Z-1596-2026
- Event ID
- 98341
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- I.T.S. GmbH
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- April 1, 2026
- Initiation Date
- January 23, 2026
- Classification Date
- March 20, 2026
- Address
- Autal 28, Lasnitzhohe, N/A, Austria
Description
Pilonplate with Angular Stability, with the below descriptions and corresponding article numbers. 1. Anterolateral Tibia Plate; Article Numbers: 21311-11, 21311-15, 21311-7, 21312-11, 21312-15, 21312-7. 2. Pilon Plate 4mm; Article Numbers: 21094-10, 21094-4, 21094-6, 21094-8. 3. Pilon Plate, Small; Article Numbers: 21093-4, 21093-6.
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
1. Article Number (Lot Numbers): 21311-11 (26/F62573), 21311-15 (26/F65048, 26/F67521), 21311-7 (26/F66165, 26/F67825), 21312-11 (26/F65049, 26/F67826), 21312-15 (26/F66166), 21312-7 (26/F65643, 26/F67827). 2. Article Number (Lot Numbers): 21094-10 (643/032304, 643/082309), 21094-4 (643/0572, 643/0571), 21094-6 (643/112202), 21094-8 (643/012309). 3.Article Number (Lot Numbers): 21093-4 (679/032311, 679/112331), 21093-6 (679/112332).
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.
101 units