FDA Enforcement Class II Ongoing

iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbers C10683 and 700-3375; in-vitro diagnostic device, urine analysis, c) iQ200 Select 1500, Part Number 700-3347, d) iQ200 Sprint, Part Number 700-3325

Recall: Z-1596-2023 · Reported May 24, 2023

Enforcement

Recall Number
Z-1596-2023
Event ID
92056
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Beckman Coulter, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
May 24, 2023
Initiation Date
April 5, 2023
Classification Date
May 17, 2023
Address
11800 Sw 147th Ave, N/A, Miami, FL, 33196-2500, United States

Description

iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbers C10683 and 700-3375; in-vitro diagnostic device, urine analysis, c) iQ200 Select 1500, Part Number 700-3347, d) iQ200 Sprint, Part Number 700-3325

Reason

Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have been investigated.

Code Info

a) iQ 200 Select, Part Numbers C10684, UDI/DI 15099590695484 and 700-3345, UD/DI 10837461001300; b) iQ 200 Elite, Part Numbers C10683, UDI/DI 15099590697129 and 700-3375, UDI/DI 10837461001355; c) iQ200 Select 1500, Part Number 700-3347, UD/DI 10837461001317; d) iQ200 Sprint, Part Number 700-3325, UDI/DI 10837461001256; All analyzers with Windows 10, Software Version 8.1;

Distribution

Worldwide Distribution

Quantity

1068 units