FDA Enforcement Class II Ongoing

FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.

Recall: Z-1595-2026 · Reported April 1, 2026

Enforcement

Recall Number
Z-1595-2026
Event ID
98341
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
I.T.S. GmbH
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
April 1, 2026
Initiation Date
January 23, 2026
Classification Date
March 20, 2026
Address
Autal 28, Lasnitzhohe, N/A, Austria

Description

FR.O.H. Calcaneus Repair System with the following description: Anterolateral Calcaneus Plate. Article Number: 21064-60.

Reason

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Code Info

Article Number (Lot Numbers): 21064-60 (679/012318, 679/0722100).

Distribution

US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Quantity

4 units