FDA Enforcement Class II Ongoing

Elecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.

Recall: Z-1595-2021 · Reported May 26, 2021

Enforcement

Recall Number
Z-1595-2021
Event ID
87705
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Roche Diagnostics Operations, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
FAX
Report Date
May 26, 2021
Initiation Date
April 7, 2021
Classification Date
May 14, 2021
Address
9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States

Description

Elecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.

Reason

Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.

Code Info

UDI: 04015630939381

Distribution

US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, ID, IL, IN,KS, KY, MD, ME, MO, NC, NE, NJ, NM, NY, OH, OK, SC, SD, TN, TX, VA, WA, WY.

Quantity

10, 489 kits