FDA Enforcement
Class II
Ongoing
Elecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.
Recall: Z-1595-2021
·
Reported May 26, 2021
Enforcement
- Recall Number
- Z-1595-2021
- Event ID
- 87705
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Roche Diagnostics Operations, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- FAX
- Report Date
- May 26, 2021
- Initiation Date
- April 7, 2021
- Classification Date
- May 14, 2021
- Address
- 9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States
Description
Elecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.
Reason
Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.
Code Info
UDI: 04015630939381
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, ID, IL, IN,KS, KY, MD, ME, MO, NC, NE, NJ, NM, NY, OH, OK, SC, SD, TN, TX, VA, WA, WY.
Quantity
10, 489 kits