Alpha Interlock Suspension System Lanyard Body: An external prosthetic device. The Alpha Lanyard includes suction with a one-way air expulsion valve plus a locking mechanism. Product design features an easy release mechanism. Available with or without a titanium pyramid. The lanyard is part of the Alpha Interlock" Suspension System. This system allows you to interchange suspension methods without remaking a socket Model, 3 part numbers - 700-AIS100, Alpha Interlock Suspension System with Alpha Lanyard - 700-AIS101, Alpha Interlock Suspension System with Alpha Lanyard w/Titanium Pyramid - 700-AIS107, Alpha Interlock Suspension System Lanyard Body ONLY
Enforcement
- Recall Number
- Z-1594-2021
- Event ID
- 87825
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- WillowWood Global Llc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- May 26, 2021
- Initiation Date
- April 20, 2021
- Classification Date
- May 14, 2021
- Address
- 15441 Scioto Darby Rd, N/A, Mount Sterling, OH, 43143-9036, United States
Description
Alpha Interlock Suspension System Lanyard Body: An external prosthetic device. The Alpha Lanyard includes suction with a one-way air expulsion valve plus a locking mechanism. Product design features an easy release mechanism. Available with or without a titanium pyramid. The lanyard is part of the Alpha Interlock" Suspension System. This system allows you to interchange suspension methods without remaking a socket Model, 3 part numbers - 700-AIS100, Alpha Interlock Suspension System with Alpha Lanyard - 700-AIS101, Alpha Interlock Suspension System with Alpha Lanyard w/Titanium Pyramid - 700-AIS107, Alpha Interlock Suspension System Lanyard Body ONLY
the firm discovered that, starting on April 1, 2020, a 3/8 dowel pin was used in place of the 5/8 pin. Over time, the shorter pin may not engage in either the upper or lower body, causing the cord to come loose or slip which could result in the loss of suspension.
Lot: 'xx' stamped on the outside of the Lanyard body manufactured form 04/01/2020 - 04/08/2021
Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IL, KS, KY, LA, MA, ME, MN, MO, MS, NJ, NM, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA and the countries of Germany, Netherlands
815 parts